Mains voltage and plug types for laboratory equipment

Most laboratory equipment is built in several electrical variants: 230 V or 115 V, 50 Hz or 60 Hz, and a choice of mains plug. The variant has to match the laboratory's supply before the order is placed. An instrument ordered in the wrong variant may run poorly, lose its warranty, or not run at all.

What to check before ordering

  • Voltage. Every country below supplies 220 to 240 V. Equipment sold for North America runs on 115 V, so the 230 V variant, usually rated 220 to 240 V, is the one to order.
  • Frequency. All the countries below run at 50 Hz except Saudi Arabia and Philippines, at 60 Hz. Motors and compressors, in centrifuges, shakers, refrigerators and ultra-low freezers, are often specified for one frequency, and running them at the other changes their speed and cooling performance.
  • Plug. Where a maker offers the instrument with several plugs, order the local one. For high-draw equipment such as autoclaves, ovens and ultra-low freezers, an adapter is not a safe substitute: the right plug, or a fixed connection by an electrician, is.
  • Supply quality. Where the grid is unstable, a voltage stabiliser or an uninterruptible power supply protects freezers and analysers. We can source both.

Voltage, frequency and plug types by country

CountryVoltageFrequencyPlug types
Morocco230 V50 HzC and E
Algeria230 V50 HzC and F
Tunisia230 V50 HzC and E
Mauritania230 V50 HzC and E
Senegal230 V50 HzC and E
Côte d'Ivoire220 to 230 V50 HzC and E
Cameroon220 to 230 V50 HzC and E
Democratic Republic of the Congo220 to 230 V50 HzC and E
Mali230 V50 HzC and E
Burkina Faso230 V50 HzC and E
Niger230 V50 HzC and E
Guinea230 V50 HzC and F
Benin230 V50 HzC and E
Togo230 V50 HzC and E
Gabon230 V50 HzC and E
Republic of the Congo230 V50 HzC and E
Chad230 V50 HzC, E and F
Nigeria230 V50 HzG
Ghana230 V50 HzG
Sierra Leone230 V50 HzG
Liberia230 V50 HzC, D, F and G
Gambia230 V50 HzG
Guinea-Bissau230 V50 HzC and F
Cabo Verde230 V50 HzC and F
Central African Republic230 V50 HzC and E
Equatorial Guinea230 V50 HzC and E
São Tomé and Príncipe230 V50 HzC and F
Lebanon230 V50 HzC and F
Syria230 V50 HzC, E, F and L
Jordan230 V50 HzC, F and G
Iraq230 V50 HzC and G
Egypt230 V50 HzC and F
Libya230 V50 HzC, F and L
Saudi Arabia230 V60 HzG
United Arab Emirates230 V50 HzG
Qatar240 V50 HzG
Kuwait230 to 240 V50 HzG
Bahrain230 V50 HzG
Oman230 to 240 V50 HzG
Sudan230 V50 HzC, D and G
Yemen230 V50 HzC, D and G
Djibouti230 V50 HzC and E
Somalia230 V50 HzC and G
Ethiopia230 V50 HzC, F and G
Kenya240 V50 HzG
Tanzania230 V50 HzG
Uganda240 V50 HzG
Rwanda230 V50 HzC, E, F and G
Burundi230 V50 HzC and E
South Sudan230 V50 HzC, D and G
Malawi230 V50 HzG
Zambia230 V50 HzG
Zimbabwe230 V50 HzG
Mozambique230 V50 HzC, F, M and N
Angola230 V50 HzC and F
Botswana230 V50 HzM
Namibia230 V50 HzC, M and N
Lesotho230 V50 HzM
Eswatini230 V50 HzC, M and N
Madagascar230 V50 HzC and E
Comoros230 V50 HzC and E
Mauritius230 V50 HzC, E and G
Seychelles240 V50 HzG
Nepal230 V50 HzC, D and M
Bangladesh230 V50 HzA, C, D, G and M
Sri Lanka230 V50 HzG
Pakistan230 V50 HzC and D
Bhutan230 V50 HzC, D and M
Maldives230 V50 HzC, D and G
Vietnam230 V50 HzA, B and C
Cambodia230 V50 HzA, B, C, F and G
Laos230 V50 HzA, B and C
Myanmar230 V50 HzA, B, C, D, E, F, G and I
Philippines230 V60 HzA, B and C
Indonesia230 V50 HzC and F
Timor-Leste230 V50 HzC, E, F and I
Mongolia230 V50 HzC and F
Papua New Guinea240 V50 HzI

Where a country uses several plug types, buildings differ. Tell us which sockets the laboratory has and we order to match. For any other country, we confirm the supply with the laboratory before ordering.