450-126.568
multi EA 4000 C with pyrolysis option
- Manufacturer
- Analytik Jena
- Reference
- 450-126.568
Specifications
- Product
- Multi EA 4000 C with pyrolysis option
- Detector
- C: NDIR
- Sample supply
- horizontal, manual or automated (accessory required)
- Precision
- At 12% C, RSD < 2 %
- Min. measurement time
- 2 min
- Parameters
- TC, TOC, TIC (accessories required), EC, OC, BOC
- Digestion mode
- High-temperature combustion, catalyst-free (oxygen), Pyrolysis (oxygen/argon)
- Gas supply
- Oxygen 2.5 Argon
- Sample quantity
- Up to 3 mg
- Limit of detection
- C: 3 ppm
- Measurement range
- C: 0 - 100 wt-%
- Dimension (width x depth x height)
- 1500 mm x 620 mm x 460 mm
- Space requirement (width x depth x height)
- 2000 mm x 700 mm x 550 mm
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- 450-126.574FPG 48 solids autosampler for multi EA 4000
- 450-900.450Liquids sampler LS 1
- 450-900.451Liquids sampler LS 2
- 450-900.452LS-T temperature controlled sampler
- 450-900.453MMS-T temperature-controlled autosampler
- 450-300.030Multi Matrix Sampler, MMS
- 450-900.461Autoinjector, type AI-EA
- 450-900.460Autoinjector, type AI-SC
- 450-900.403Autoinjector, without syringe
- 450-300.013Automatic Boat Drive, ABD with flame sensor technology
- 450-126.420Gas sampling system, GSS for multi EA 5000 / 5100
- 450-126.410Gas sampling system for compEAct / multi EA 5010