DIANA

Diana, atmospheric distillation analyser

Manufacturer
Anton Paar
Reference
DIANA

Specifications

Model
Diana
Type
Atmospheric distillation analyser
Standards
ASTM D1078, ASTM D4530, ASTM D524, ASTM D850, ASTM D86, EN 60751, IP 123, IP 195, ISO 17025, ISO 3405
Standard test methods fuel distillation
ASTM D86 (groups 0 to 4), EN ISO 3405 (groups 0 to 4), IP 123, JIS K 2254, GOST 2177
Solvent distillation
ASTM D1078, ASTM D850, IP 195, JIS K 0066
Additional methods
ASTM D524, ASTM D4530, Benzinum Medicinale
Measurement vapour temperature sensor
Pt100, metal or glass, Class A (DIN EN 60751)
Vapour temperature range
0 °C to 450 °C
Vapour temperature calibration
Integrated sensor memory for up to 12 calibration points; automatic probe ID detection
Calibration certificate
Factory calibration certificate included (room temperature, 100 °C, 180 °C); ISO 17025 accredited calibration available optionally
Distillation rate
Up to 10 mL/min, programmable
Volume measurement
Optical static detection system with contact image sensor (CIS) and anti-condensation technology
Volume range
0 mL to 103.5 mL
Volume accuracy
±0.1 mL

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item