9040-0208

CB 260, CO2 incubator with hot air sterilisation and heat sterilisable CO2 sensor, with O2 control 0.2 to 20%, 267 L interior, +6 °C above ambient temperature to 50 °C, humidity 90 to 95 % RH, 200 to 230 V 50/60 Hz, designation CB260-230V-O

Manufacturer
BINDER
Reference
9040-0208
Product line
CO2 incubators
Item type
CO2 incubator

Specifications

Designation
CB260-230V-O
Option model
With O 2 control 0.2 to 20%
Temperature range
+6 °C above ambient temperature to 50 °C
Temperature uniformity at 37°C
0.4 ± K
Temperature fluctuation at 37°C
0.1 ± K
Recovery time after door was opened for 30 s at 37°C
6 min
Humidity range
90 ...95 % RH
CO₂ range
0 ...20 Vol.-% CO₂
CO₂ measuring technology
IR
CO₂ recovery time after door was opened for 30 s at 5 vol. % CO₂
5 min
Standard O₂ control range
0.2 ...20 Vol.-% O₂
O₂ control ranges with option: O₂ range
10 ...90 Vol.-% O₂
O₂ recovery time after door was opened for 30 s at 5 vol. % O₂
15 min
Rated voltage
200...230 V
Power frequency
50/60 Hz
Nominal power
1,5 kW
Unit fuse
10,0 A
Phase (nominal voltage)
1~
Interior volume
267 L
Net weight of the unit (empty)
125 kg
Load per rack
10 kg
Permitted load
40 kg
Wall clearance back
100 mm
Wall clearance sidewise
50 mm
Width net
740 mm
Height net
1,020 mm
Depth net
785 mm
Interior width
620 mm
Interior height
750 mm
Interior depth
575 mm
Inner doors
1
Unit doors
1
Sound-pressure level
41 dB(A)
Energy consumption at 37°C
65 Wh/h
Number of shelves (std./max.)
3/8

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line