Dissolution On-/Offline System UV/Vis
Automated sampling with UV-Vis analysis and sample storage
- Manufacturer
- Erweka
- Reference
- Dissolution On-/Offline System UV/Vis
- Product line
- Dissolution Testers
- Item type
- Dissolution tester
Specifications
- Weight
- 42 kg
- Dimensions (H x W x D)
- 850 x 650 x 650 mm
- Voltage
- 115/230 V; 50/60 Hz
- Speed
- 20-250 U/min
- Vessel volume
- 400 ml / 1000 ml / 2000 ml
- Interfaces
- 1x RS-232, 2x USB, 2x Ethernet/RJ45
- Test stations (DT 956)
- 6 in 2 rows
- Test stations (DT 957)
- 7 in 2 rows
- Test stations (DT 958)
- 8 in 2 rows
- USP methods
- USP 1 / USP 2 / USP 5 / USP 6
- Fuses
- 2 A
- Protection class
- I/EN 61140
- Protection type
- IP 21/IEC 529
- Operation
- Touchscreen 7", 800x480 Pixel
- Sampling positions
- High-head / Low-head / Cleaning mode
- Ambient temperature during operation
- +10 °C to +30 °C (ambient temperature min. -5 °C below set temperature)
- Storage & transport temp.
- +5 °C to +40 °C
- Relative humidity
- 25-80 % non condensing
- Test stations (DT 9512)
- 12 in 2 rows
- Test stations (DT 9513)
- 13 in 2 rows
- Test stations (DT 9514)
- 14 in 2 rows
- Channels
- 6,7 or 8
- Max. capacity (not for 25 ml test tubes)
- 26 sample intervals
- Valve
- Integrated 3 way valves
- FRL racks
- 26 x 8 / 1.5 ml HPLC vials, 26 x 8 / 4.0 ml HPLC vials, 26 x 8 / 10 ml test tubes, 18 x 8 / 25 ml test tubes
- Voltage +- 10%
- 115 - 250 VAC; 50 / 60 Hz
- Pump type
- PVP 1220/1420 (for DT 9510)
- Valves
- /
- Accuracy
- +/- 0.5 ml
- System compatibility
- DT Online System, DT Offline System, DT On-/Offline System
- Media replacement
- Standard
- Double filtration (optional)
- Only when first filtration with poroplast filters. No media replacement possible when double filtration.
- Required type of sample collector
- FRL 654 / 754 / 854
- Advantages
- Basic pump possible with DT 950/9510, needs regular replacement of tubing
- Filter requirements
- Pore size examples: 0.45 μm, with other sizes on request
- Supported filters
- - PALL Membrane Filter ACRODISC - Whatman Roby 25 syringe filters for robotic systems
- Stations
- 12 valves for 6 stations / 16 valves for 8 stations
- Optical design
- Double beam spectrophotometer with fixed spectral bandwidth (Specord 210 with variable spectral bandwidth)
- Spectral bandwidth
- 1.4 nm
- Light source
- Combination of deuterium and halogen lamp
- Detector
- Two silicon photodiodes
- Spectrometric system
- Monochromator with imaging grating and aspherical quartz-coated optics
- Baseline deviation
- ± 0.0005 A (200–1000 nm; Slit 1.4 nm)
- Zero point transmission
- ± 0.05 %T (200–1000 nm; Slit 1.4 nm)
- Wavelength range
- 190–1100 nm
- Wavelength accuracy
- ± 0.1 nm (Deuterium line at 656.1 nm)
- Wavelength reproducibility
- ≤ 0.02 nm
- Wavelength registration speed
- Up to 12000 nm/min
- Wavelength Min. data interval
- 0.02 nm
- Photometric range
- -3 to 3 A
- Photometric accuracy UV
- ± 0.010 A
- Photometric accuracy vis
- ± 0.003 A
- Photometric reproducibility
- ≤ 0.0005 A
- Stray light 198 nm (KCl)
- ≤ 0.3 %T
- Stray light 220 nm (nal)
- ≤ 0.03 %T
- Stray light 240 nm (nal)
- ≤ 0.03 %T
- Stray light 340 nm (NaNO2)
- ≤ 0.02 %T
- Baseline noise at 500 nm (RMS)
- ≤ 0.0001 A
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- DT 950Digital dissolution tester, USP 1, 2, 5 and 6, 6 to 8 test stations, TestAssist
- DT 9510Digital dissolution tester, USP 1, 2, 5 and 6, 12 to 14 test stations, TestAssist
- DT LightDissolution tester for USP 1, 2, 5 and 6, 6 or 8 test stations, fixed drive head
- Dissolution Offline SystemDT 950 or DT 9510 with automated sampling and sample storage
- Dissolution On-/Offline System HPLCSemi-automatic dissolution system with direct injection to HPLC and sample storage
- RoboDis II+Fully automated dissolution system for 10, 20 or 40 test runs, piston pump
- BioDis RRT 10Reciprocating cylinder dissolution tester, USP 3 and 7
- DFZ IIFlow-through cell dissolution tester, USP 4
- DRTChewing gum dissolution tester
- MediPrepMedia preparation for 8 and 16 L of dissolution medium, MediPrep 820 and 1620
- Disso.NET 4Dissolution software, 21 CFR Part 11 compliant