Dissolution On-/Offline System UV/Vis

Automated sampling with UV-Vis analysis and sample storage

Manufacturer
Erweka
Reference
Dissolution On-/Offline System UV/Vis

Specifications

Weight
42 kg
Dimensions (H x W x D)
850 x 650 x 650 mm
Voltage
115/230 V; 50/60 Hz
Speed
20-250 U/min
Vessel volume
400 ml / 1000 ml / 2000 ml
Interfaces
1x RS-232, 2x USB, 2x Ethernet/RJ45
Test stations (DT 956)
6 in 2 rows
Test stations (DT 957)
7 in 2 rows
Test stations (DT 958)
8 in 2 rows
USP methods
USP 1 / USP 2 / USP 5 / USP 6
Fuses
2 A
Protection class
I/EN 61140
Protection type
IP 21/IEC 529
Operation
Touchscreen 7", 800x480 Pixel
Sampling positions
High-head / Low-head / Cleaning mode
Ambient temperature during operation
+10 °C to +30 °C (ambient temperature min. -5 °C below set temperature)
Storage & transport temp.
+5 °C to +40 °C
Relative humidity
25-80 % non condensing
Test stations (DT 9512)
12 in 2 rows
Test stations (DT 9513)
13 in 2 rows
Test stations (DT 9514)
14 in 2 rows
Channels
6,7 or 8
Max. capacity (not for 25 ml test tubes)
26 sample intervals
Valve
Integrated 3 way valves
FRL racks
26 x 8 / 1.5 ml HPLC vials, 26 x 8 / 4.0 ml HPLC vials, 26 x 8 / 10 ml test tubes, 18 x 8 / 25 ml test tubes
Voltage +- 10%
115 - 250 VAC; 50 / 60 Hz
Pump type
PVP 1220/1420 (for DT 9510)
Valves
/
Accuracy
+/- 0.5 ml
System compatibility
DT Online System, DT Offline System, DT On-/Offline System
Media replacement
Standard
Double filtration (optional)
Only when first filtration with poroplast filters. No media replacement possible when double filtration.
Required type of sample collector
FRL 654 / 754 / 854
Advantages
Basic pump possible with DT 950/9510, needs regular replacement of tubing
Filter requirements
Pore size examples: 0.45 μm, with other sizes on request
Supported filters
- PALL Membrane Filter ACRODISC - Whatman Roby 25 syringe filters for robotic systems
Stations
12 valves for 6 stations / 16 valves for 8 stations
Optical design
Double beam spectrophotometer with fixed spectral bandwidth (Specord 210 with variable spectral bandwidth)
Spectral bandwidth
1.4 nm
Light source
Combination of deuterium and halogen lamp
Detector
Two silicon photodiodes
Spectrometric system
Monochromator with imaging grating and aspherical quartz-coated optics
Baseline deviation
± 0.0005 A (200–1000 nm; Slit 1.4 nm)
Zero point transmission
± 0.05 %T (200–1000 nm; Slit 1.4 nm)
Wavelength range
190–1100 nm
Wavelength accuracy
± 0.1 nm (Deuterium line at 656.1 nm)
Wavelength reproducibility
≤ 0.02 nm
Wavelength registration speed
Up to 12000 nm/min
Wavelength Min. data interval
0.02 nm
Photometric range
-3 to 3 A
Photometric accuracy UV
± 0.010 A
Photometric accuracy vis
± 0.003 A
Photometric reproducibility
≤ 0.0005 A
Stray light 198 nm (KCl)
≤ 0.3 %T
Stray light 220 nm (nal)
≤ 0.03 %T
Stray light 240 nm (nal)
≤ 0.03 %T
Stray light 340 nm (NaNO2)
≤ 0.02 %T
Baseline noise at 500 nm (RMS)
≤ 0.0001 A