RTS-8 Plus

Personal multi-channel bioreactor with non-invasive real-time OD, pH and pO2 measurement

Manufacturer
Grant Instruments
Reference
RTS-8 Plus

Specifications

Specification
Details
Light source
Laser
Measurement wavelength (λ)
850 ± 15 nm
Measurement range
0–100 OD600
E. coli factory calibration measurement range
0–50 OD600
S. cerevisiae factory calibration measurement range
0–75 OD600
User calibration error (0.1–6 OD600)
±0.3
User calibration error (6–50 OD600)
≤5%
User calibration error (50–75 OD600)
≤10%
Measurement periodicity per hour
1–60
Temperature setting range
+15°C to +60°C
Temperature control range
+15°C below ambient to +60°C
Temperature stability
±0.3°C
Sample temperature accuracy (20–37°C)
±1°C
Tube sockets
8
Sample working volume range
3–50 ml
Speed control range
150–2700 rpm
Reverse spin time @150–250 rpm
0 s
Reverse spin time @250–300 rpm
2–60 s
Reverse spin time @300–2700 rpm
0–60 s
Display
LCD
Minimum PC requirements
Intel/AMD Processor, 1 GB RAM, Windows Vista/7/8/8.1/10/11, USB 2.0
Dimensions (W×D×H)
350 × 690 × 300 mm
Weight
20 kg
Nominal operating voltage
AC 230 V, 50 Hz
Power consumption
500 W (3.15 A)
O₂ sensor
+
Range
0–100%
Accuracy
±0.05% O₂ at 0.2%; ±0.4% O₂ at 20.9%
Drift
<0.03% O₂ within 30 days
Temperature range
Up to 40°C
Response time (t90)
6 s
Storage stability
18 months
pH sensor
+

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line