L002501

PRO RP 250 EC, -50 to 200 °C, filling volume 4.4 L, 208 to 220 V 60 Hz, CEE7/7 plug

Manufacturer
LAUDA
Reference
L002501

Specifications

Model
PRO RP 250 EC
Working temperature range
-50 to 200 °C
Operating temperature range
-50 to 200 °C
Ambient temperature range
5 to 40 °C
Temperature stability
0.05 ± K
Heater power max.
2.3 kW
max. power consumption
3.5 kW
Current consumption
16 A
max. discharge pressure
0.7 bar
Pump suction max.
0.4 bar
max. pump flow pressure
22 L/min
Pump_flow_suction
20 L/min
In / outlet connection thread (outside)
M16 x 1
In / outlet Ø olives
13 mm
Min. filling volume
2.4 L
max. filling volume
4.4 L
Dimensions (WxDxH)
300 x 430 x 675 mm
Weight
48 kg
Refrigerant stage
R-452A (GWP 2140); 0.300 kg; 0.6 t CO2-eq
Power supply
208-220 V; 60 Hz
Power plug
Power cord with angled plug (CEE7/7)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line