L002674

Integral IN 250 XTW, -45 to 220 °C, filling volume 8.7 L, 230 V 50 Hz, CEE7/7 plug, cooling water flow rate 4.2 L/min

Manufacturer
LAUDA
Reference
L002674

Specifications

Model
Integral IN 250 XTW
Working temperature range
-45 to 220 °C
Ambient temperature range
5 to 40 °C
Temperature stability
0.05 ± K
Heater power max.
3.5 kW
max. power consumption
3.7 kW
Current consumption
16 A
max. discharge pressure
3.1 bar
max. pump flow pressure
65 L/min
In / outlet connection thread (outside)
M30 x 1,5
Inlet/outlet hose size
3/4″
Min. filling volume
2.5 L
max. filling volume
8.7 L
Water cooling connection thread (outside)
3/4 ″
Recommended cooling water temperature
15 °C
Cooling water temperature max
30 °C
Cooling water flow rate
4.2 L/min
Recommended pressure difference cooling water
4 bar
Pressure difference cooling water min.
0.8 bar
Pressure difference cooling water max.
5 bar
Maximum pressure cooling water
10 bar
Dimensions (WxDxH)
430 x 550 x 760 mm
Weight
106 kg
Noise level
57 dB(A)
Refrigerant stage
R-449A (GWP 1397); 0.450 kg; 0.6 t CO2-eq
Power supply
230 V; 50 Hz
Power plug
Power cord with angled plug (CEE7/7)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line