L003308

Integral IN 2560 PW, -40 to 140 °C, filling volume 48.1 L, 400 V 3/PE 50 Hz and 460 V 3/PE 60 Hz

Manufacturer
LAUDA
Reference
L003308

Specifications

Model
Integral IN 2560 PW
Working temperature range
-40 to 140 °C
max. pressurized overlay
4 bar
Ambient temperature range
5 to 40 °C
Temperature stability
0.1 ± K
Heater power max.
24 kW
max. power consumption
37 kW
Current consumption
60 A
max. discharge pressure
6.0 bar
max. pump flow pressure
100 L/min
In / outlet connection thread (outside)
M38 x 1,5
Inlet/outlet hose size
1″
Min. filling volume
12.1 L
max. filling volume
48.1 L
Water cooling connection thread (outside)
3/4 ″
Recommended cooling water temperature
15 °C
Cooling water temperature max
30 °C
Cooling water flow rate
29 L/min
Recommended pressure difference cooling water
3 bar
Pressure difference cooling water min.
0.8 bar
Pressure difference cooling water max.
10 bar
Maximum pressure cooling water
10 bar
Dimensions (WxDxH)
1100 x 895 x 1865 mm
Weight
625 kg
Noise level
74 dB(A)
Refrigerant stage
R-449A (GWP 1397); 4.000 kg; 5.6 t CO2-eq
Power supply
400 V; 3/PE; 50 Hz & 460 V; 3/PE; 60 Hz
Power plug
Electric connection box without power cord, without plug