UTRID-16

LogTag UTRID-16, multi-trip USB PDF temperature data logger

Manufacturer
LogTag
Reference
UTRID-16

Specifications

Sensor measurement range
-30°C to +60°C (-22°F to +140°F).
Operating temperature range
-30°C to +65°C (-22°F to +149°F).*
Storage temperature range
0°C to +40°C (32°F to +104°F).
Rated temperature reading accuracy
Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F). Better than ±0.7°C (±1.3°F) for -25°C to -20°C (-13°F to -4°F) & +40°C to +60°C (+104°F to +140°F). Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration.
Rated temperature reading resolution
0.1°C/F. LogTag Analyser currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag.
Sensor reaction time
Typically less than 7 minutes (T90) in moving air (1m/s) in EN12830:1999.
Recording capacity
16,129 temperature readings. 56 days @ 5min logging, 168 days @ 15min logging. Statistics memory (for display on LCD):max Min values and alarm duration values.
Sampling interval
Configurable from 30 seconds to 18 hours.
Logging start options
Push button start (with configurable start delay from 1 minute to 72 hours) or specific date & time.
Recording indication
“REC” symbol on LCD.
Download time
Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report. Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag Analyser (if configured).
Environmental
IEC 60529: IP64 with USB cap fitted.
Power source
UTRID-16R: CR2032 3V LiMnO2 Battery (Replaceable). UTRID-16: CR2450 3V Li-MnO2 UN38.3 approved coin cell
Battery life
UTRID-16R: 1 year storage, 1 year of normal use (based on 15 minute logging, download data monthly). The battery low symbol will appear on the LCD if the UTRID-16 has a low battery.UTRID-16: 1 year of storage, then 2 years of operation with normal use (6 minute logging …
Real time clock
Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Connection interface
USB 2.0, A-type plug.
Software
PDF Reader, LogTag Analyser 2.8 or higher.
Multi-language support
Multi-language support for English, Spanish, French, Russian and German.
Size
93mm(H) x 54.5mm(W) x 8.6mm(T) including protective USB cap.
Weight
39g.
Case material
Polycarbonate.
Footnote
*(-30°C to +65 °C applies to Loggers with serial numbers 6x45x15811 or greater. Loggers with serial numbers prior to this have an operating temperature range of -25°C to +60°C).

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line