Model 26-1
Ludlum Model 26-1, pancake GM detector, stainless steel screen (79% open)
- Manufacturer
- Ludlum Measurements
- Reference
- Model 26-1
- Product line
- Contamination monitors and friskers
Specifications
- Product
- Ludlum Model 26-1, pancake GM detector, stainless steel screen (79% open)
- Ludlum part number
- m26-1-series
- Detector
- Pancake GM detector, stainless steel screen (79% open)
- Window Area
- Active: 15.51 cm 2 (2.4 in 2 ) Open: 12.26 cm 2 (1.9 in 2 )
- Efficiency
- Alpha: 239 Pu - 11% Beta: 99 Tc - 14%; 32 P - 32%; 14 C - 2%; 90 Sr/ 90 Y - 22%; 125 I - 0.2% Gamma: 5.5 cps per μSv/hr (3300 cpm per mR/hr) [ 137 Cs]; 99m Tc - ≤ 1%
- Energy Range
- 16.3 keV to 13. MeV
- Display
- 3½ digit LCD with large 12.7 mm (0.5 in.) digits, low-battery indicator, MAX, ALARM Units: (k)cps, (k)cpm, (k)dpm, (k)Bq, mR/hr, µSv/h
- Display Range
- 0.00 cps to 19.9 kcps 0.00 cpm to 999 kcpm 0.00 Bq to 19.9 kBq 0.00 dpm to 999 kdpm 0.00 to 500 mR/h .000 to 1999 µSv/h
- Linearity
- ±10%
- Backlight
- Built-in ambient light sensor automatically activates low-power LED backlight, or may be configured for 'Continuous On' (will reduce battery life)
- Controls
- ON / OFF / QUIET: Press to turn ON; Tap to alternate between click audio and QUIET; Hold for OFF MODE: Alternates between RATE (count rate), MAX (captures peak rate), and COUNT (preset count time from 0 to 20 min.) UNITS
- Response Time
- User-selectable from 1 to 60 seconds, or auto-response rate FAST or SLOW
- Resolution Time
- Approximately 110 µs as defined by IEC 60325
- Alarms
- Count rate and scaler alarm setpoints adjustable over the display range
- Overload Protection
- High count rate saturation protection prevents false display of lower count rates, indicated by flashing display and audio alarm
- Loss of Count Protection
- After 60 seconds of no pulses from detector, unit will flash a zero reading and the alarm audio will be triggered
- Click Audio
- Greater than 60 dB at 0.6 m (2 ft)
- HV Testpoint
- Accessible by removing battery cover, allows HV measurement and pulse injection for calibration using optional cable ( see Options )
- Temperature Range
- -20 to 50 °C (-4 to 122 °F) May be certified for operation from -40 to 65 °C (-40 to 150 °F)
- Environmental Rating
- NEMA 3, IP 53
- Power
- Two “AA” batteries
- Battery Life
- Approximately 1000 hours of operations (as low as 500 hours with backlight configured for 'Continuous On'), 16-hour low battery warning
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 26Ludlum Model 26, integrated frisker
- Model 26-2Ludlum Model 26-2, integrated frisker with timed frisk
- Model 26-3Ludlum Model 26-3, high range frisker
- Model 26SLudlum Model 26S, integrated scintillator frisker
- Model 239-1FLudlum Model 239-1F, floor contamination monitor
- Model 240Ludlum Model 240, alpha-beta floor and wall monitor
- Model 3002Ludlum Model 3002, /fm series alpha-beta floor monitor