Model 3000
Ludlum Model 3000, survey meter
- Manufacturer
- Ludlum Measurements
- Reference
- Model 3000
- Product line
- Survey meters, Geiger counters and ratemeters
Specifications
- Product
- Ludlum Model 3000, survey meter
- Ludlum part number
- 48-4035
- Compatible Detectors
- GM, scintillation, or proportional
- Display
- 3 digit auto-ranging LCD with large 20 mm (0.8 in.) digits. Units: (k)cps, (k)cpm, (k)Bq, (k)dpm, (µ)(m)R(/h), (µ)(m)Sv(/h). Indicators: bar graph, MAX, ALARM, USB, audio, low battery.
- Display Range
- 0.0 cps to 99.9 kcps 0.00 cpm to 999 kcpm 0.00 Bq to 99.9 kBq 0.00 dpm to 999 kdpm 0.00 µR/h to 999 R/h 0.00 µSv/h to 999 Sv/h Operational display ranges are dependent on the detector and dead time configuration. The min
- Linearity
- Readings within 10% of true value
- Backlight
- Built-in ambient light sensor automatically activates low-power LED backlight, unless internal dipswitch is set to continuous-On (will reduce battery life)
- Controls
- ON/OFF: Extra-long press to turn ON; Short press to acknowledge alarms; Extra-long press plus 3 seconds to turn OFF UNITS: Short press changes the units between count rate, dose/exposure, or disintegration; Extra-long pr
- Data Logging
- Can store up to 1000 data points internally. Requires Lumic Data Logger Kit, see Options.
- Response Time
- User-selectable from 1 to 60 seconds, or auto-response rate FAST or SLOW
- Alarms
- Two count rate, exposure/dose, and count alarm setpoints adjustable over the display range
- Overload Protection
- High count rate saturation protection prevents false display of lower count rates. “OL” will appear on the display and an audio warning will be triggered.
- Loss of Count Protection
- After a user-defined time of no pulses from detector, typically 60 seconds, the unit will flash a zero reading, and an audio alarm will be triggered.
- Overrange
- If reading exceeds the predefined detector range, the instrument will flash a maximum reading and trigger an audio warning.
- Dead Time Correction
- Employs first and second order corrections for extended performance
- High Voltage
- 400 to 1500 Vdc
- Threshold
- -4 to -100 mVdc
- Audio
- Approximately 4.5 kHz; Click audio greater than 65 dB at 0.6 m (2 ft); Alarm audio greater than 72 dB
- Power
- Four alkaline “AA” batteries
- Battery Life
- Approximately 750 hours of operation (as low as 100 hours with backlight configured for continuous-on), 16-hour low battery warning
- Construction
- High-impact plastic with water-resistant rubber seals and separate battery compartment
- Temperature Range
- -20 to 50 °C (-5 to 122 °F) May be certified for operation from -40 to 65 °C (-40 to 150 °F)
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 3005Ludlum Model 3005, digital survey meter with internal detector
- Model 5Ludlum Model 5, Geiger counter
- Model 6Ludlum Model 6, Geiger counter
- Model 2401-ECLudlum Model 2401-EC, pocket-size survey meter
- Model 2401-EC2Ludlum Model 2401-EC2, pocket-size survey meter
- Model 2401-EC2ALudlum Model 2401-EC2A, pocket-size survey meter with alarm
- Model 2401-ECALudlum Model 2401-ECA, pocket-size survey meter with alarm
- Model 2401-EWLudlum Model 2401-EW, end-window GM pocket-size survey meter
- Model 2401-EWALudlum Model 2401-EWA, end-window GM pocket-size survey meter with alarm
- Model 2401-PLudlum Model 2401-P, pancake GM pocket-size survey meter
- Model 2402Ludlum Model 2402, pocket-size survey meter with alarm
- Model 2403Ludlum Model 2403, pocket-size survey meter