Model 3001

Ludlum Model 3001, multi-detector survey meter

Manufacturer
Ludlum Measurements
Reference
Model 3001

Specifications

Product
Ludlum Model 3001, multi-detector survey meter
Ludlum part number
48-4036
Compatible Detectors
GM, scintillation, or proportional
Display
3-digit auto-ranging LCD with large 20 mm (0.8 in.) digits. Units: (k)cps, (k)cpm, (k)Bq, (k)dpm, (μ)(m)R(/h), (μ)(m)Sv(/h). Indicators: bar graph, MAX, ALARM, USB, audio, low battery.
Display Range
0.0 cps to 99.9 kcps 0.00 cpm to 999 kcpm 0.00 Bq to 99.9 kBq 0.00 dpm to 999 kdpm 0.00 μR/h to 999 R/h 0.00 μSv/h to 999 Sv/h Operational display ranges are dependent on the detector and dead time configuration. The min
Linearity
Within 10% of true value
Backlight
Built-in ambient light sensor automatically activates LED backlight, unless internal dipswitch is set to continuous-On (will reduce battery life)
User Controls
ON/ACK: Extra-long press to turn ON, short press to adjust audio level and acknowledge alarms, long press to reset mode, extra-long press plus 3 seconds to turn OFF UNITS: Short press changes the units between count rate
Data Logging
Can store up to 1000 data points internally. Requires Lumic Data Logger Kit, see Options.
Response Time
User-selectable from 1 to 60 s, or Auto-Response Rate FAST or SLOW
Alarms
Two adjustable count rate, exposure/dose, and scaler alarm setpoints adjustable over the display range
Overload Protection
High count rate saturation protection prevents false display of lower count rates. "OL" will appear on the display and an audio warning will be triggered.
Loss of Count Protection
After a user-settable time interval (default 60 seconds) of no pulses from detector, unit will flash a zero reading and the alarm audio will be triggered
Dead Time Correction
Employs first and second order corrections for extended performance
Overrange
If the reading should exceed the predefined detector range, the instrument will flash a maximum reading and trigger an audio warning.
High Voltage
400 to 1500 Vdc
Threshold
-2 to -100 mVdc
Audio
Approximately 4.5 kHz, click audio greater than 65 dB at 0.6 m (2 ft), alarm audio greater than 72 dB
Power
Four alkaline “AA” batteries
Battery Life
Approximately 500 hours (as low as 60 hours with backlight configured for continuous-on), 16-hour low battery warning
Construction
High-impact plastic with water-resistant rubber seals and separate battery compartment
Temperature Range
-20 to 50 °C (-5 to 122 °F) May be certified for operation from -40 to 65 °C (-40 to 150 °F)