Model 3003

Ludlum Model 3003, multi-detector survey meter

Manufacturer
Ludlum Measurements
Reference
Model 3003

Specifications

Product
Ludlum Model 3003, multi-detector survey meter
Ludlum part number
3003-series
Compatible Detectors
Up to four external GM, scintillation, proportional, or neutron detectors
Internal Detector (3003i Series only)
Energy compensated GM or scintillation detector (see Configurations tab for options)
High Voltage
300 to 2400 Vdc
Linearity
Readings within 10% of true value
Modes
Four operational modes including rate, max, integrated dose, and count
Display
3-digit auto-ranging LCD with large 20 mm (0.8 in.) digits. Units: (k)(M)c, (k)cps, (k)(M)cpm, (k)(M)d, (k)(M)Bq, (k)(M)Bq/cm 2 , (k)(M)dpm, (µ)(m)(k)(M)R, (µ)(m)(k)R/h, (µ)(m)(k)(M)Sv, (µ)(m)Sv/h, (µ)(m)(k)(M)Rem, (µ)(m
Display Range
00.0 cps to 999 kcps 00.0 cpm to 99.9 Mcpm 00.0 Bq to 999 MBq 00.0 Bq/cm 2 to 999 MBq/cm 2 00.0 dpm to 999 Mdpm 0.00 µR to 999 MR 0.00 µR/h to 9.99 kR/h 0.00 µSv to 999 MSv 0.00 µSv/h to 999 Sv/h 0.00 µRem to 999 MRem 0.
Backlight
Built-in ambient light sensor automatically activates LED backlight unless internal dipswitch is set to continuous-on
User Controls
ON/AUD: Extra-long press to turn ON; Short press to adjust audio level; Extra-long press plus 3 seconds to turn OFF UNITS: Short press Changes the units between count rate, dose/exposure, or activity DETECTOR: Short pres
Response Time
User-selectable from 1 to 60 seconds, auto SLOW, auto FAST, fixed SLOW (22 s), fixed FAST (4 s)
Alarms
Three count rate, exposure/dose, count, and accumulated dose alarm set-points adjustable over the display range
Self-Diagnostics
Detects and reports internal instrument issues, indicated by graphic FAULT display and audio warning
Overload Protection
High count rate saturation protection prevents false display of lower count rates. OVERLOAD will appear on the display and an audio warning will be triggered.
Loss of Count Protection
After a user-defined time, typically 60 seconds, with no pulses from detector, the unit will flash a zero reading, an audio alarm will be triggered, and LOSS OF COUNTS will appear on the display
Overrange
If reading should exceed the predefined detector range, the unit will flash a maximum reading, an audio warning will be triggered, and OVERRANGE will appear on the display
Audio
"Click" audio greater than 65 dB at 0.6 m (2 ft.), Alarm audio greater than 72 dB at 0.6 m (2 ft.), approximately 4.5 kHz
Calibration Protection
Internal dipswitch to disable front-panel calibration parameter changes
Power
Four "AA" batteries
Battery Life
Approximately 175 hours, 16-hour low battery warning
Construction
High-impact polycarbonate with rubber seals and separate battery compartment

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line