Model 375-9
Ludlum Model 375-9, digital area monitor
- Manufacturer
- Ludlum Measurements
- Reference
- Model 375-9
- Product line
- Area radiation monitors
Specifications
- Product
- Ludlum Model 375-9, digital area monitor
- Ludlum part number
- 48-3560
- Detector
- Type: external Model 45-9 Ion Chamber Detector Housing: aluminum with beige powder-coat finish , 7.6 x 25.7 cm (3 x 10.1 in.) [Dia. x L] Ion Chamber: 17 cm 3 volume, stainless steel wall chamber filled with xenon at 1.01
- Display
- 4-digit LED with 2 cm (0.8 in.) character height
- Operating Range
- Low (default): 0.1 to 999.9 mR/hr or 1 to 9999 μSv/h Mid: 1 to 9999 mR/hr or 0.01 to 99.99 mSv/h High: 0.01 to 99.99 R/hr or 0.1 to 999.9 mSv/h Selection of range and units must be decided at time of purchase.
- Linearity
- Readings with 10% of true value with detector connected
- Status Indicators
- STATUS: green light, indicates the instrument is functioning properly LOW ALARM: yellow light and slow beep (1 per second), can be set at any point in range HIGH ALARM: red light and fast beep (4 per second), can be set
- Calibration Controls
- Accessible from the front of the instrument (protective cover provided)
- Over-Range
- Display reading of “– – – –” and activated low and high alarms indicate that the radiation field being measured has exceeded the counting range of the instrument (or when dead time correction accounts for more than 75% o
- Audio
- Intensity can vary from approx. 68 to 100 dB through operation of the external rotary baffle and the internal voltage connection. Frequency is approx. 3 kHz.
- Data Output
- 9-pin connector providing 5-decade logarithmic output, RS‑232 output, signal ground connection, FAIL and ALARM signals (current sink), and direct connection to battery and ground
- RS-232 Output
- 2-second dump for computer data logging
- Power
- 95 to 135 Vac (178 to 240 Vac available), 50 to 60 Hz single phase (less than 100 mA), 6-volt sealed lead-acid rechargeable battery (built in)
- Battery Life
- Typically 48 hours in non-alarm condition, 12 hours in alarm condition
- Battery Charger
- Battery is continuously trickle-charged when instrument is connected to line power and turned on
- Construction
- Aluminum housing with ivory powder coat finish
- Temperature Range
- -20 to 50 °C (-4 to 122 °F) May be certified for operation from -40 to 65 °C (-40 to 150 °F)
- Size (H x W x L)
- 18.7 x 24.6 x 6.4 cm (7.4 x 9.7 x 2.5 in.), electronics only
- Weight
- 2.9 kg (6.5 lb), excluding detector
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this item