Fluorophos

Fluorophos fluorometric alkaline phosphatase tester for milk and dairy products

Manufacturer
Nova Biomedical
Reference
Fluorophos

Specifications

Product
Fluorophos fluorometric alkaline phosphatase tester for milk and dairy products
Electrical power requirement
100 to 240 VAC (50/60 Hz)
Fuses (2)
250V time deley (type T): 1 Amp
Power consumption
180 Watts
Memory backup
Integral lithium cell; 10 year life (typical); not user replaceable
Sample test volume
See test kit instruction
Aspirated sample volume
150μL
Sample capacity
Single sample
Readout
20 Character vacuum florescent display
Unit
FLU (florescence Unit)
Result Unit
See test kit instructions
Zeroing
Automatic
Resolution
1 FLU
Drift
Less Than 3 FLU per hour
Cuvette size
12 X 75 mm round
Warmup time
Cuvette chamber 15 minutes, Heating block 30 minutes
Test time
3 minutes
Optical
90 Optical bench with light source, fixed filter, detector
Light Source
Light Emitting diode
Filter
Excitation 440 nm bandpass, Emission 550 nm longpass
Storage temperature
-40 to 113° F/-40 to 45°C