Autopol V
Digital polarimeter, six wavelengths 365 to 633 nm, TempTrol 15 to 35 °C, 21 CFR Part 11
- Manufacturer
- Rudolph Research Analytical
- Reference
- Autopol V
- Product line
- Polarimeters
- Item type
- Polarimeter
Specifications
- Measuring mode
- Optical Rotation Specific Rotation Specific Rotation Plus Concentration User Defined Scale Sugar Degrees °Z(ISS)
- Measuring scale
- Degrees Arc, % Concentration
- Measuring range
- ± 89.99° Arc Optical Rotation ± 999.99°Arc Specific Rotation 0-99.9% Concentration
- Resolution
- 0.001° Arc Optical Rotation 0.001% Concentration 0.001 Specific Rotation
- Reproducibility
- 0.002° Arc
- Standard model accuracy
- 0.002° up to 1°, 0.2% up to 5°, 0.01° above 5°
- AP model specs
- Accuracy (546nm and 589nm): ±0.002° Arc over ±89.9° Arc Accuracy for other wavelengths is the same as the standard model Selectable Resolution 0.01°, 0.001°, 0.0001° Arc Optical Rotation
- Prism
- Glan Thomson calcite quartz
- Optical wavelengths
- 365nm, 405nm, 436nm, 546nm, 589nm, 633nm (other wavelengths available)
- Wavelength selection
- Touchscreen Selectable: 6 standard
- TempTrol range
- Automatic Electronic Heating & Cooling 15°-35°C (Standard) ±0.2° C (Standard)
- TempTrol accuracy
- ±0.2°C
- Temp. probe range
- 10°-100°C
- Temp. probe accuracy
- ±0.1°C
- Acid resistance
- Hastelloy measurement cell and Silco Steel sample chamber (optional)
- Measurement time
- 4°/sec slewing rate & 5 sec nominal settling time
- Light source
- Tungsten-halogen 6V, 20W, avg. 2,000 hr life
- Sample chamber
- Accepts sample tubes up to 200mm
- Data storage
- 32 GB Non-removable Compact Flast
- Communication interface
- 3 - USB Ports 2 - RS232 Ports Ethernet Port for Network Connection
- Calibration
- Automatic calibration by push-button
- Operating system
- Windows 10 IOT ENTERPRISE LTSC
- Display
- 8 inch colour LCD with LED backlight providing 800 x 600 resolution with 400 nits of luminance
- User interface
- Touchscreen
- Automatic sensitivity control
- Measures samples with transmittance as low as 0.01 % (up to O.D. 4.0)
- Input power
- 100-240 VAC, 50/60 Hz
- Operating dimensions
- 30"W x 17"H x 10.5"D, 762mm W x 432mm H x 267mm D
- Shipping dimensions
- 30"W x 11.5"H x 17"D, 762mm W x 431.8mm H x 279.4mm D
- Operating weight
- 85 lbs (39 kg)
- Shipping weight
- 120 lbs. (54kg)
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Autopol IAutomatic polarimeter, 589 nm fixed wavelength, accuracy 0.01° arc, built-in temperature probe, TempTrol option
- Autopol IIAutomatic polarimeter, 589 and 546 nm, optional 365 to 633 nm, accuracy 0.01° arc, TempTrol option
- Autopol IIIAutomatic polarimeter, dual wavelength 589 and 546 nm, accuracy 0.002° arc up to 1°, resolution 0.001°
- Autopol IVAutomatic polarimeter in one, two or six wavelength versions, 365 to 633 nm, resolution 0.001°
- Autopol V PlusPharmaceutical polarimeter, six wavelengths, TempTrol 15 to 35 °C, 21 CFR Part 11, Hastelloy cell for 6 M HCl
- Autopol VIPolarimeter, six wavelengths 365 to 633 nm, resolution 0.0001°, accuracy 0.0003°, TempTrol 15 to 35 °C