5495307H0--FF--V

Sartorius Sartopore Platinum Midicaps 0.2 µm Size 0 pack of 2, filter size size 0 (0.52 m² / 5.6 ft²), sterilization method Autoclavable, Inlet: ¾" Sanitary Outlet: ¾" Sanitary

Manufacturer
Sartorius
Reference
5495307H0--FF--V

Specifications

Product type
Process filtration: sterilising-grade cartridges, capsules and integrity testers
Application
Single Use Processing
Industry
Pharma / Biotech
Modality
Form & Fill Buffer & Auxiliaries Filtration
Scale
Production | Small Scale
Bacterial Retention
Quantitative Retention of 1 × E7 CFU/cm² Brevundimonas Diminuta ATCC 19146 per ASTM methodology
Bacterial Endotoxins
< 0.25 EU/mL as determined by the LAL test
Biological Reactivity
All materials of this filter element meet the requirements of the current USP Biological Reactivity tests <88> for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).
Extractables
The total amount of extractables is well below the limits established by the current USP unter "Sterile Water for Injection".
Connector Inlet
3/4" Tri-clamp 25 mm
Connector Outlet
3/4" Tri-clamp 25 mm
Connector(s)
Inlet: ¾" Sanitary | Outlet: ¾" Sanitary
Filter Size
Size 0 (0.52 m² / 5.6 ft²)
Filtration Area
1 m²
Outer Diameter
108.5 mm (77 mm without Connection / Valves)
Capsule/Cartridge Format
Midicaps
Filter Format
Capsules
Maximum Allowable Back Pressure
2 bar (29 psi) at 20°C; 1.5 bar (21.8 psi) at 50°C
Maximum Diffusion
14 mL/min at 2.5 bar | 36 psi
Minimum Required Bubble Point
3.5 bar | 51 psi
Vent Valve
Standard
Filter Material
Polyethersulfone (PES)
Housing Material
Polypropylene (PP)
Material o-ring | seal
Silicone (SI)
Membrane Material
Polyethersulfone (PES)
Support Fleece
Polypropylene (PP)
Maximum Differential Pressure
20 °C: 5 bar, 50 °C: 3 bar
Maximum Operating Pressure
5 bar (72.5 psi)
Non-Fiber Releasing
This filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
Pore Size Final-filter
0.2 µm
Pore Size Pre-filter
0.45 µm
Delivery Condition
Sterile - Autoclaved
Sterilization Method
Autoclavable
Sterilization Procedures
Autoclaving: at 134°C, 30 min, 2 bar, min. 25 Cycles
Application
Adeno-associated Virus
Product family
Sartopore Platinum

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line