TOC-4200

Online TOC analyser, 680 °C combustion catalytic oxidation with NDIR, 0 to 5 up to 0 to 1,000 mgC/L full scale, up to six streams

Manufacturer
Shimadzu
Reference
TOC-4200
Product line
TOC analyzers
Item type
TOC analyser

Specifications

Measurement items
TC, NPOC Optional: IC, POC, TOC (= TC-IC, = NPOC + POC), TN
Measurement principle
680℃ combustion catalytic oxidation - NDIR detection method
Measurement range
From 0 - 5 to 0 - 1,000 mgC/L f.s. (0 to 20,000 mgC/L f.s. with dilution function)
Repeatability
Within ±2 % f.s. *1
Span stability
Within ±2 % f.s. /day (temperature variation within 5℃) *1
Zero stability
Within ±2 % f.s. /day (temperature variation within 5℃) *1
Measurement cycle
4 minutes min. (NPOC, residual IC 2 % max.) *2
Sample injection method
Syringe pump/sliding injection port
IC removal method
Acidify/sparge within syringe
Sample dilution function
Dilution within syringe; dilution factor 2 to 50 Dilution accuracy: Within ±2 % (x2 to x20); within ±5 % (x21 to x50)
Stream switching function
Streams 1 to 6 can be switched (with the relevant option)
Display, operation
Colour LCD with touch screen
Conversion function
Conversion with a linear expression using required indices (Conversion expression must be determined in advance.)
Load calculation function
Using the analog input board (option) permits load calculations. (Non-isolated analog inputs; load resistance: 100 ohms)
Data storage function
Store 20,000 past on-line measured values (equivalent to one year's data at 30-minute measuring cycle). Offers trend graph displays.
Data storage device
Store measured values or measurement conditions to USB memory *3 . (USB 1.1 or 2.0 USB memory; FAT16 or FAT32; no encryption)
Recorder
Recording width 100 mm, 6-dot recorder (optional)
Printer
Thermal printer, 40 characters per line; 110 mm chart width (optional)
Analog output
Selectable between 4 – 20 mA DC or 0 – 16 mA DC (isolated outputs) With attached option: Up to 12 outputs (load resistance: 500Ω max.)
Contact output signal
Alarm signals Major failure, instrument alarm, power cut, CPU alarm, concentration alarm (upper limit/lower limit/upper upper limit/lower lower limit), measurement halted Event signals Maintenance, measurement ready, on-line operation, measuring sample, calibrating, regenerating catalyst, measuring control sample …
Contact input
Start calibration, stop measurement, start on-line measurement, reset alarm, measurement start flow line 1 to 6, halt measurement (output contact capacity: 1.4 V DC min., 10 mA min.)
Communication function
Compatible with RS-485, RS-232C (option), and Modbus Use the optional Web monitoring unit set to browse measured values and alarms on a Web browser over a LAN.
Carrier gas
Pressurized air, high-purity air, oxygen *4 ; supply pressure: 250 to 300 kPa Optional: High-purity nitrogen *4
Power supply
AC 100 to 240 V ±10%, 10 A, 50/60 Hz
Ambient temperature
Within 1 to 40℃
Construction
Indoor wall mounting (optional stand available)
Dimensions
W550 x D383 x H1240 mm (excluding protrusions) (W22 x D15 x H49 inches)
Weight
Approx. 70 kg (154 lbs.)
Measurement items (*Optional)
NPOC, TC, TOC (TC-IC)* TOC (NPOC+POC)*

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line