Specifications

Product
Setaflash Series 3 ActiveCool Flash Point Tester
UN class 3 Non-viscous flammable liquids
Small Scale Flash Point Testing
EPA 1020 A & B
Setaflash Closed-Cup Method for determining ignitability
CLP regulations EC no 1272/2008
European Regulation (EC) No 1272/2008 on classification, labelling and packaging of substances and mixtures
SW-846 1020C
Standard Test Methods for Flash Point by Setaflash (Small Scale) Closed-Cup Apparatus
Temperature range
10 to 135 °C
Cup material
Aluminium or stainless steel insert
Test modes
Rapid Equilibrium and Ramp
Sample size
2 ml for flash points up to 100 °C 4 ml for flash points above 100 °C
Test duration
1 minute below 100 °C, 2 minutes above 100 °C, or user determination 1 to 99 minutes
Ramp rate
2 °C/min ramp
Heating and cooling system
Peltier Cell
Power
200 W (maximum)
Voltage
100/240 V, 50/60 Hz
Gas supply
Laboratory gas, 3 kPa (0.44 psi) maximum pressure or Butane from optional Portable Gas Tank 13667-0
Size (HxWxD)
19.5 cm x 29.5 cm x 14 cm
Weight
3 kg
Methods
ASTM D3278, ISO 3680 (Obs), ASTM E502, ISO 3679, ASTM D3828, IP 523, IP 524, ASTM D7236, IP 534, ASTM D1655