LSTAR1125

Orion Lab Star EC112 Conductivity Bench Meters, Orion Lab Star EC112 conductivity/salinity/TDS bench meter standard kit, For a Wide Variety of Conductivity Samples in The 1 μS/cm to 20 mS/cm Range

Manufacturer
Thermo Fisher Scientific
Reference
LSTAR1125

Specifications

Fisher Scientific catalogue number
17701859
Includes
011050MD 2-cell K = 1.0 epoxy-body conductivity/ATC probe; 011007 1413 μS conductivity standard 5 x 60 mL; meter with electrode stand, PC cable, universal power adapter
Accuracy (conductivity)
±0.5 % of the reading ± 1 LSD
Outputs
Computer or printer (STARA-106)
For use with (application)
For a Wide Variety of Conductivity Samples in The 1 μS/cm to 20 mS/cm Range
Measurement modes
Conductivity with Temperature, Salinity with Temperature, TDS with Temperature
Compatible probes
2-cell and 4-cell conductivity electrodes
Cell constant
0.100 to 10.00 cm-1
Calibration
Conductivity, salinity, TDS and temperature
Calibration points (conductivity)
1-5 points conductivity
Calibration points
1 point salinity, 1-5 points TDS, 1 point temperature
Range (conductivity)
0.00 μS to 500.0 mS
Resolution (conductivity)
0.01, 0.1, 1 μS; 0.01, 0.1 mS
Temperature compensation
Linear (0 to 10 %/°C) or off
Temperature coefficient
0 to 10 %/°C
Temperature (metric) reference
15.0, 20.0, 25.0 or 30.0°C
Range (salinity)
0.00 to 80.0 ppt; 0.00 to 42.0 PSU
Resolution (salinity)
0.01, 0.1 ppt; 0.01, 0.1 PSU
Accuracy (salinity)
±0.5 % of the reading ± 1 LSD
Salinity measurement modes
Natural Seawater (ppt) or Practical Salinity (PSU)
Range (TDS)
0.00 ppm to 500.0 ppt
Resolution (TDS)
0.01, 0.1, 1 ppm; 0.01, 0.1 ppt
Accuracy (TDS)
±0.5 % of the reading ± 1 LSD
TDS factor range
0.40 to 1.00
Temperature range
-5.0°C to +105.0°C
Resolution (temperature)
0.1°C, 0.1°F
Accuracy (temperature)
±0.3°C, ±0.5°F
Temperature selection
Manual or automatic with ATC temperature probe
Temperature probe calibration
1 point temperature offset calibration
Display type
5 in. backlit LCD
Backlight option
Yes
Memory
Non-volatile
Data points
500 Sets with Date and Time Stamp
Log function type
Automatic with Auto-Read and Timed Read Modes, Manual with Continuous Read Mode
Data management details
Continuous, Auto-Read and Timed read modes
Alarm output
Calibration reminder alarm
Inputs
8-pin MiniDIN for conductivity with built-in ATC
Certifications/Compliance
CE, TUV 3-1, FCC Class A
IP rating
IP54
AC adapter
Universal 100-240 V 50/60 Hz power supply with twist-and-lock meter connection and interchangeable wall plugs for US/Japan, Euro, UK/Singapore, Australia/New Zealand and China outlets
Dimensions (L x W x H)
187.2 x 155.2 x 61.7 mm, 7.37 x 6.11 x 2.43 in.
Weight
700 grams
Warranty
3 year meter warranty
Type
Conductivity Meter Kit

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line