VF-D11-A-01
Vanquish Diode Array Detectors and Multiple Wavelength Detectors, 250 Hz Max., Vanquish Diode Array Detector FG, UV-vis Detection (Up to 10 Channels), < 1 mAU/hour at 254 nm, 100–240 V AC, 50/60 Hz
- Manufacturer
- Thermo Fisher Scientific
- Reference
- VF-D11-A-01
- Product line
- HPLC Detectors
- Item type
- HPLC and UHPLC system
Specifications
- Fisher Scientific catalogue number
- 17719869
- Data collection rate
- 250 Hz Max.
- Detection mode
- UV-vis Detection (Up to 10 Channels)
- Detector type
- Single-Beam
- Type
- Diode Array Detector
- Drift
- < 1 mAU/hour at 254 nm
- Flow cell volume
- Standard, path length 10 mm (13 μL, SST) Standard, biocompatible, path length 10 mm (13 μL, PEEK) Semi-analytical, path length 7 mm (5 μL, SST) Semi-micro, pathlength 7 mm (2.5 μL, SST) Semi-micro, biocompatible, path length 7 mm (2.5 μL, titanium) Diagnostic cell
- GLP features
- Predictive performance functions: lamp age and ignitions, lamp intensity degradation, leak detection, service monitoring period. All system parameters logged in the Chromeleon CDS Audit Trail.
- Lamp
- Deuterium lamp, tungsten lamp
- Linearity
- < 5% at 2.2 AU (typically < 5% at 2.7 AU)
- Max. cell operating pressure
- Standard biocompatible: 5 MPa (50 bar, 720 psi) Other flow cells: 12 MPa (120 bar, 1740 psi)
- Noise
- <± 6 μAU at 254 nm (standard flow cell, bandwidth: 4 nm, slit width: wide, time constant: 2 sec)
- No. of channels
- 10 + 3D field
- PC connection
- USB 2.0; 3-port HUB to connect further Vanquish modules
- Photodiode array
- 1024 element
- Safety features
- Power-up diagnostics of optics, cooling fans, motors and electronics. Leak detection and safe leak handling.
- Slit width
- Settable: Wide, narrow
- Spectral bandwidth
- Pixel resolution: 0.6 nm (average)
- Temperature range
- 5 to 35°C; 20 to 80% RH (non-condensing)
- Validation functions
- Internal validation with holmium oxide filter
- Wavelength accuracy
- ± 1 nm
- Wavelength range
- 190–800 nm
- Wavelength repeatability
- ± 0.1 nm
- Weight
- 16.3 kg
- Electrical requirements
- 100–240 V AC, 50/60 Hz, max. 245 W / 255 VA
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.