VLBLATD0

Varioskan LUX Multimode Microplate Reader, Measurement Technology: Absorbance, Fluorescence Intensity, Luminescence, Time-resolved Fluorescence, AlphaScreen, Reading Options: Top and Bottom, Dispensers: None

Manufacturer
Thermo Fisher Scientific
Reference
VLBLATD0

Specifications

Fisher Scientific catalogue number
15360757
Compatibility
Robot
Detector type
Photomultiplier Tube (PMT)
Dimensions (L x W x H)
58 x 53 x 51 cm (23 x 21 x 20 in.)
Interface
PC Software (Thermo Scientific SkanIt Software)
Incubator temperature
Ambient +4° to 45°C
Light source
Xenon Flash Lamp, LED
Measurement speed
15 sec. for 96-well Plate, 45 sec. for 384-well Plate, 135 sec. for 1536-Well Plate (Minimum Times)
Measurement technology
Absorbance, Fluorescence Intensity, Luminescence, Time-Resolved Fluorescence, AlphaScreen
Plate type
6 to 1536-well Plates (Absorbance 6 to 384-well Plates), μDrop Plates
Product line
Varioskan
Reading options
Top, Bottom
Shaking
Orbital
Spectral scanning speed
<2.2 sec./Well 400 to 500 nm, 1 Flash, 2 nm Step
Read-out range
0 to 6 Abs
Wavelength selection
Dual Monochromators (Abs), Quad Monochromators (FL); 2 Ex, 2 Em, Filters (Lum., TRF, Alpha)
Software type
SkanIt Software
Accuracy
0.003 Abs or ±2%, at 200 to 399 nm (0 to 2 Abs), 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs)
Accuracy (photometric)
0.003 Abs or ±2% at 200 to 399 nm (0 to 2 Abs), 0.003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs)
Depth
58 cm
Dynamic range
Fluorescence intensity: top reading > 6 decades, bottom reading > 5.5 decades, Luminescence: > 7 decades
Emission filters
Available for purchase
Excitation filters
Fixed filters (Alpha Screen/AlphaLISA excitation wavelength 680 nm; TRF 334 nm)
Height
51 cm
Includes
SkanIt PC Software, Research Edition
Linearity
96-well plate: ±2% (0 to 4 Abs at 450 nm), 384-well plate: ±2% (0 to 3 Abs at 450 nm)
Measurement type
Endpoint, Kinetic, Spectra, Multipoint, Kinetic Spectra
Precision
SD < 0.001 Abs or CV < 0.5%, at 450 nm (0 to 3 Abs)
Sensitivity
Fluorescence intensity (top reading): <0.4 fmol fluorescein/well (black 384-well plate), Fluorescence intensity (bottom reading): <4 fmol fluorescein/well, (clear bottom black 384 square well plate) …
Software language versions
English, German, French, Spanish, Portuguese, Italian, Japansese, Chinese
Wavelength range
Absorbance: 200-1000 nm; Fluorescence intensity: Excitation: 200-822 nm, Emission: 270-840 nm: TRF: Excitation 334 nm, Emission 400 to 700 nm; Luminescence: 360-670 nm; AlphaScreen: Excitation 680 nm, Emission 400-660 nm
Weight
57 kg
Width
53 cm

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item