SRIC-4

LogTag SRIC-4, single-use temperature data logger

Manufacturer
LogTag
Reference
SRIC-4

Specifications

Sensor measurement range
-30°C to +60°C (-22°F to +140°F).
Operating temperature range
-30°C to +60°C (-22°F to +140°F).
Storage temperature range
0°C to +40°C (32°F to +104°F).
Rated temperature reading accuracy
Better than ±0.5°C (±0.9°F) for -5°C to +30°C (23°F to 86°F). ±0.8°C (±1.5°F) or better for measurements in other areas of rated range. Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration.
Rated temperature reading resolution
Less than 0.1°C/F. LogTag Analyser currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag.
Sensor reaction time
Typically less than 5 minutes (T90) in moving air (1m/s).
Recording capacity
3968 temperature readings. 13 days @ 5min logging, 41 days @ 15min logging.
Sampling interval
Configurable from 1 minute to 18 hours.
Logging start options
Push button start or specific date & time.
Recording indication
Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator.
Download time
Typically less than 4 seconds for full memory (3968 readings), depending on computer or readout device used.
Alarms
1 configurable upper and 1 configurable lower alarms.
Environmental
IP65 (roughly equivalent to NEMA 4).
Power source
CR2032 3V LiMnO2 Battery (Fixed).
Battery life
20 months storage, followed by 90 days normal use.
Real time clock
Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Size
77mm(H) x 54.5mm(W) x 8.6mm(T).
Weight
28g.
Case material
Polycarbonate.

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item