TRID30-7

LogTag TRID30-7, multi-trip LCD 30-day temperature data logger

Manufacturer
LogTag
Reference
TRID30-7

Specifications

Sensor measurement range
-30°C to +60°C (-22°F to +140°F).
Operating temperature range
-30°C to +60°C (-22°F to +140°F).
Storage temperature range
-10°C to +40°C (14°F to +104°F).
Rated temperature reading accuracy
Better than ±0.5°C (±0.9°F) for measurements from -20°C to +40°C (-4°F to +104°F), typically ±0.3°C (0.6°F). Better than ±0.8°C (±1.5°F) for other measurements – typically ±0.5°C (0.9°F). Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration.
Rated temperature reading resolution
0.1°C (0.2°F) for measurements -40°C to +50°C (-40°F to +122°F). 0.2°C (0.4°F) for measurements +50°C to +70°C (+122°F to +158°F). LogTag Analyser currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag.
Sensor reaction time
Typically less than 5 minutes (T90) in moving air (1m/s) in EN12830:1999.
Recording capacity
7,650 temperature readings. 53 days @ 10min logging, 80 days @ 15min logging. Day summary statistics memory (for display on LCD): up for 30 days of max Min/duration and alarm trigger statistics.
Sampling interval
Configurable from 30 seconds to hourly.
Logging start options
Push button start or specific date & time. Optional start delay of up to 18 hours.
Recording indication
State indicator “REC”
Download time
Typically less than 5 seconds for full memory (7,650 readings), depending on computer or readout device used.
Environmental
IP65 (roughly equivalent to NEMA 4) when vertically mounted or hung.
Power source
TRID30-7R (Replaceable Battery): CR2032 3V LiMnO2 TRID30-7F (Fixed Battery): CR2450 3V LiMnO2.
Battery life
TRID30-7R (Replaceable Battery): 1 year storage, followed by 1 year of operation with normal use (6 minute logging, statistics reviewed on the display no more than once daily for no longer than 30 seconds each time, download data monthly). TRID30-7F (Fixed Battery): 1 year storage …
Real time clock
Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Connection interface
Interface Cradle
Software
LogTag Analyser
Size
93mm(H) x 54.5mm(W) x 8.6mm(T).
Weight
TRID30-7R: 41g. TRID30-7F: 43g.
Case material
Polycarbonate.

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item