Model 177

Ludlum Model 177, benchtop counter

Manufacturer
Ludlum Measurements
Reference
Model 177

Specifications

Product
Ludlum Model 177, benchtop counter
Ludlum part number
48-1632
Indicated Use
Multi-purpose radiation monitoring, surveying, frisking
Compatible Detectors
GM, Scintillation
Linearity
Within 5% of full scale
Meter Dial
0 - 500 cpm, 0 - 1.5 kV, BAT TEST
Meter
6.4 x 6.4 cm (2.5 x 2.5 in.) 1 mA, DC movement
Multipliers
x1, x10, x100, x1k
Response
Toggle switch control selects FAST (2.2 s) or SLOW (22 s) response, for 90% of full scale reading
Alarm Set
Front-panel locking control allowing alarm to be set at any point on the meter (can be read on meter)
Reset
Pushbutton to zero meter and/or reset alarm
Calibration Controls
Accessible from back of instrument (protective cover provided)
Alarm
Indicated by red lamp and audible tone (greater than 60 dB at 61 cm [2 ft])
High Voltage
Adjustable from 400 to 1500 V
Input Sensitivity
Adjustable from -10 to -100 mV
Audio
Built-in unimorph speaker, with volume control, to provide click-per-event audio or audible alarm indicator
Data Out
9-pin series "D" connector allowing for external connections of the following items: Battery Unregulated supply Instrument common ground Alarm signal (current sink) Pulse out External background subtract Recorder output
Recorder Output
Correlated to meter reading (adjustable to 0 to 1.25 V at 1 mA)
Power
95 to 250 Vac and 6 V sealed lead-acid rechargeable battery (built-in)
Battery Life
Typically 50 hours in non-alarm condition (battery condition can be checked on meter)
Battery Charger
Battery is continuously trickle-charged when instrument is connected to line power and turned on
Size (H x W x L)
12.7 x 20.3 x 15.2 cm (5 x 8 x 6 in.)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item