Model 3276

Ludlum Model 3276, multipurpose digital meter

Manufacturer
Ludlum Measurements
Reference
Model 3276

Specifications

Product
Ludlum Model 3276, multipurpose digital meter
Ludlum part number
48-4160
Compatible Detectors
GM, Scintillation, Proportional
LCD Display
3-digit auto-ranging LCD with large 20 mm (0.8 in.) digits. Units: (k)cps, (k)cpm, (k)Bq, (k)dpm, (µ)(m)R(/h), (µ)(m)Sv(/h). Indicators: bar graph, MAX, ALARM, USB, audio, low battery.
Display Range
0.00 cps to 999 kcps 0.00 cpm to 999 kcpm 0.00 Bq to 999 kBq 0.00 dpm to 999 kdpm 0.00 µR/h to 999 R/h 0.00 µSv/h to 999 Sv/h NOTE : Operational display ranges are dependent on the detector and dead time configuration. M
Linearity
Readings within 10% of true value
Backlight
Continuously on to provide better visibility
User Controls
ON/OFF: Extra-long press to turn ON; Short press to acknowledge alarms; Extra-long press plus 3 seconds to turn OFF UNITS: Short press changes the units from count rate, dose/exposure, or disintegration, an extra-long pr
Response Time
User-selectable from 1 to 60 seconds, or Auto-Response Rate FAST or SLOW
Alarms
Two count rate, exposure/dose, and count alarm setpoints adjustable over the display range
Overload Protection
High count rate saturation protection prevents false display of lower count rates. “OL” will appear on the display and an audio warning will be triggered.
Loss of Count Protection
After a user-settable time interval (default 60 seconds) of no pulses from detector, the unit will flash a zero reading and the alarm audio will be triggered
Dead Time Correction
Employs first and second order corrections for extended performance
Overrange
If the reading should exceed the predefined detector range, the instrument will flash a maximum reading and trigger an audio warning.
High Voltage
550 Vdc
Threshold
-4 to -100 mV, typically set to -35 mV for GM detectors
Audio
Approximately 4.5 kHz, click audio greater than 65 dB at 0.6 m (2 ft), alarm audio greater than 72 dB
Power
Four alkaline “AA” batteries for portable use, or 9 Vdc input via wall mount transformer (input 100 - 250 Vac, 50 - 60 Hz).
Battery Life
Approximately 50 hours with the backlight ON, 16-hour low-battery warning
Construction
Aluminum housing with powder coat finish and subsurface printed front panel
Temperature Range
-20 to 50 °C (-5 to 122 °F) May be certified for operation from -40 to 65 °C (-40 to 150 °F)
Environmental Rating
NEMA rating of 2 or IP rating of 51

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item