Specifications

Fisher Scientific catalogue number
11621489
AC adapter
Universal 50-60 Hz, 100-240 VAC power adapter (sold separately, Cat. No. 1010003)
Outputs
USB, RS232
Accuracy (barometer)
±6mmHg
Accuracy (conductivity)
0.5% of reading ±1 digit > 3 μS; 0.5% of reading ±0.01 μS ≤ 3 μS
Accuracy (ISE)
±0.2 mV or ±0.05% of reading whichever is greater
Accuracy (mV)
±0.2 mV or ±0.05% of reading whichever is greater
Accuracy (pH)
±0.002
Accuracy (polarographic DO)
±0.2mg/L or ±2% of reading, whichever is greater
Accuracy (RDO)
±0.1mg/L up to 8mg/L, ±0.2mg/L from 8 to 20mg/L, ±10% of reading up to 50mg/L
Accuracy (resistivity)
0.5% of reading ±1 digit
Accuracy (salinity)
0.5% of reading ±1 digit
Accuracy (TDS)
0.5% of reading ±1 digit
Advanced ISE functions
Segmented (point-to-point) slope, non-linear selectable auto-blank, low concentration range stabilit
Alarm output
High/low limit alarm, calibration due alarm
Backlight option
Yes (Selectable)
Barometric pressure correction
400 to 850mmHg, automatic using built-in barometer or manual entry
Battery life
800 hr.
Battery type
4 AA (included)
Calibration
PH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
Calibration modes
Water-saturated air, air-saturated water, manual (Winkler titration), zero point
Calibration points
1 point relative mV (RmV) mode, 1 point ORP mode
Calibration points (conductivity)
Up to 5
Calibration points (ISE)
1 to 5
Calibration points (pH)
1 to 5
Cell constant
0.001 to 199.9cm-1
Certifications/Compliance
CE, TUV 3-1, FCC Class A
Data management details
Transfer single, range or all data points to printer or computer
Data points
5000 with Date and Time Stamp
Channels
3
Display type
LCD Graphic, Backlit
IP rating
IP67
Measurement modes
pH, mV, Relative mV, TDS, ORP with Temperature, Salinity, Conductivity, Resistivity with Temperature
Percent saturation range (polarographic DO)
0.0 to 600.0 % saturation
Range (conductivity)
0.001 μS/cm to 3000 mS/cm
Range (mV)
±2000.0 mV
Range (pH)
-2 to 20
Range (resistivity)
2 ohm to 100.0 meg-ohm
Range (salinity)
0.06 to 80.00 psu, 0.05 to 42.00 ppt
Range (TDS)
0.001 ppm to 200.0 ppt
Resolution (ISE)
0.0001 minimum, 1 to 3 significant digits (user selectable)
Resolution (pH)
0.1/0.01/0.001
Temperature range
-5.0°C to +105.0°C
Includes
Star A329 pH/ISE/Conductivity/RDO/DO Meter with Batteries; 8107UWMMD ROSS Gel-filled Epoxy pH/ATC Triode with 3 Meter Cable; 013010MD DuraProbe Epoxy 4-Probe (K=0.475) Conductivity Probe with 3 Meter Cable; 087010MD RDO Optical DO Probe with 3 Meter Cable
Inputs
BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), 9 pin MiniDIN (RDO optical or polarographic DO sensor with built-in temperature)
Isopotential point
7.000pH
Keypad
Comprehensive with menu-specific function keys and dual purpose scroll/shortcut keys
Languages
English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
Log function type
Automatic data logging with Auto-Read and Timed measure modes; manual data logging with Continuous m
Memory
Non-volatile memory preserves data log, calibration log and meter settings
Methods
10 per channel with password protection
Model
Orion Star A329 pH/ISE/mV/RmV/ORP/Conductivity/TDS/Salinity/Resistivity/RDO Optical/Dissolved Oxygen/Temperature Portable Meter
Percent saturation relative accuracy (RDO)
±2 % saturation ≤ 200 % saturation, ±10 % saturation > 200 % saturation
Percent saturation resolution (polarographic DO)
0.1, 1 % saturation
Percent saturation range (RDO)
0.0 to 500.0 % saturation
Percent saturation relative accuracy (polarographic DO)
±2 % saturation or ±2% of reading whichever is greater
Percent saturation resolution (RDO)
0.1, 1 % saturation
Power supply
4 AA batteries (included) or universal AC adapter (sold separately)
Pressure range
400 to 850mmHg
Pressure resolution
1mmHg
Product type
Portable Meter Kit
Range (ISE)
0.0001 to 19900 with selectable units of ppm, Molar, mg/L, %, ppb or none
Range (ORP)
±2000.0 mV
Range (polarographic DO)
0.00 to 90.00mg/L
Range (RDO)
0.00 to 50.00mg/L
Range (relative mV)
±2000.0 mV
Relative humidity range
5 to 85 %, non-condensing
Resolution (conductivity)
0.001 μS minimum, auto ranging up to 4 significant digits
Resolution (mV)
0.1mV
Resolution (polarographic DO)
0.01, 0.1mg/L
Resolution (RDO)
0.01, 0.1mg/L
Resolution (resistivity)
1 ohm or 0.1 meg-ohm, auto ranging
Resolution (salinity)
0.01 psu or 0.01 ppt minimum, auto ranging
Resolution (TDS)
0.001 ppm minimum, auto ranging up to 4 significant digits
Salinity factor
0.0 to 45.0 ppt, automatic correction using conductivity reading or manual entry of sample salinity
Salinity measurement modes
Practical salinity (psu) or natural sea water (ppt)
Stability
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting
TDS factor range
Linear (0.02 to 9.99)
Temperature accuracy
±0.1°C
Temperature compensation
Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off
Temperature (english) operating
41°F to 113°F (ambient)
Temperature (metric) operating
5°C to 45°C (ambient)
Temperature (metric) reference
5, 10, 15, 20, 25°C
Temperature probe calibration
1 point temperature offset calibration
Temperature resolution
0.1°C, 0.1°F
Temperature selection
Manual or automatic with ATC temperature probe
Warranty
3 Year Meter
Dimensions (L x W x H)
24cm x 10cm x 6.5cm (9.5 in. x 3.9 in. x 2.6 in.)