START9300

Orion Star T930 ion titrator without electrode

Manufacturer
Thermo Fisher Scientific
Reference
START9300
Product line
Benchtop Titrators

Specifications

Fisher Scientific catalogue number
15892097
Accuracy (ISE)
±0.2 mV or ±0.05% of reading, whichever is greater
Accuracy (mV)
±0.2 mV
Burette functions
Automatic rinse cycles for flushing and discrete dispense volume with continuous option; Exceeds acc
Functions available
Sample titration, titrant standardization, calibration and direct measure
Accuracy (temperature)
±0.2 °C
Advanced ISE functions
Multiple known addition (MKA); automated calibration and calculation of sample concentration by adding aliquots of standard to sample using the burette
Alarm output
Notifications for Titration Cycle Complete, Maximum Titrant Volume, Data Log Full, Calibration Due, Maintenance Reminder
Backlight option
Yes, adjustable
Blank options
Fixed value or variable value using titration
Burette sizes
10 mL, 20 mL (included) or 50 mL
Calibration
Ion concentration (ISE) and temperature
Calibration points (ISE)
1 to 5
Certifications/Compliance
CE, TUV 3-in-1, FCC, EN/EIC61326-1, IEC 61010
Channels
1
Cycles per titration run
1 to 5 cycles, option to exclude cycle results from average results and RSD calculations
Data management details
Data log export as CSV or PDF file, short or long format
Data points
100 data sets each: sample titration, titrant standardization, calibration and direct measure
Display type
5.7 in. colour capacitive touchscreen, 640 x 480 resolution
Type
Ion Titrator
Equivalence points
1 to 2
For use with (application)
Ion concentration titrations including salt in food products, calcium in juice and milk, total hardness of drinking water and wastewater, chloride in drinking water and wastewater, ammonia and Total Kjeldahl Nitrogen (TKN) of wastewater and surfactants in consumer products
Includes
20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, literature on USB drive, 110-240V power supplies
Inputs
BNC (ion selective electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), stirrer input (stirrer probe)
IP rating
IP51
Languages
Chinese, English, French, German, Italian, Japanese, Korean, Portuguese and Spanish
Material
ABS Plastic
Memory
Non-volatile memory preserves data logs and settings
Methods
Up to 10 methods with optional password protection; import/export methods via USB flash drive
Model
Orion Star T930
Outputs
3 USB ports for printer, USB flash drive and computer software connections
Precision
0.5% RSD, dependent on environment and handling conditions
Preset endpoints
1 to 3
Probe type
Ion selective electrode (ISE), reference half-cell electrode, ATC temperature probe, stirrer probe
Range (ISE)
0.0001 to 19990
Range (mV)
-2000.0 to +2000.0 mV
Relative humidity range
5 to 85%, non-condensing
Resolution (ISE)
0.0001 minimum, 1 to 4 significant digits (user selectable)
Resolution (mV)
0.1 mV
Resolution (temperature)
0.1 °C
Sample ID
Auto-incremental, manual or off
Stirrer probe speeds
5 user selectable speeds (250 to 3700 RPM)
Temperature (english) operating
41°F to 104°F
Temperature (metric) operating
5°C to 40°C
Temperature probe calibration
Optional 1 point offset calibration
Temperature range
-5.0°C to +100.0°C
Temperature selection
Manual or automatic
Time and date
Yes, with nonvolatile battery backup
Titrant determination
Standardization titration or manual concentration input
Titration process control
Routine, quick, careful or user defined
Titration setup wizard
Yes, using Start a New Titration workflow
Titration technique
Equivalence point or preset endpoint
Titration types
Direct or back titration
Updateable firmware
Yes, using USB flash drive
Warranty
1 year replacement warranty
Dimensions (L x W x H)
10 x 16 x 14 in. (25.4 x 40.6 x 35.6 cm)
Electrical requirements
100-240V, 50/60 Hz

Add this reference to my quote