START9301
Orion Star T930 ion titrator salt content kit
- Manufacturer
- Thermo Fisher Scientific
- Reference
- START9301
- Product line
- Benchtop Titrators
Specifications
- Fisher Scientific catalogue number
- 15802107
- Accuracy (ISE)
- ±0.2 mV or ±0.05% of reading, whichever is greater
- Accuracy (mV)
- ±0.2 mV
- Burette functions
- Automatic rinse cycles for flushing and discrete dispense volume with continuous option; Exceeds acc
- Functions available
- Sample titration, titrant standardization, calibration and direct measure
- Accuracy (temperature)
- ±0.2 °C
- Advanced ISE functions
- Multiple known addition (MKA); automated calibration and calculation of sample concentration by adding aliquots of standard to sample using the burette
- Alarm output
- Notifications for Titration Cycle Complete, Maximum Titrant Volume, Data Log Full, Calibration Due, Maintenance Reminder
- Backlight option
- Yes, adjustable
- Blank options
- Fixed value or variable value using titration
- Burette sizes
- 10 mL, 20 mL (included) or 50 mL
- Calibration
- Ion concentration (ISE) and temperature
- Calibration points (ISE)
- 1 to 5
- Certifications/Compliance
- CE, TUV 3-in-1, FCC, EN/EIC61326-1, IEC 61010
- Channels
- 1
- Cycles per titration run
- 1 to 5 cycles, option to exclude cycle results from average results and RSD calculations
- Data management details
- Data log export as CSV or PDF file, short or long format
- Data points
- 100 data sets each: sample titration, titrant standardization, calibration and direct measure
- Display type
- 5.7 in. colour capacitive touchscreen, 640 x 480 resolution
- Type
- Ion Titrator Salt Content Kit
- Equivalence points
- 1 to 2
- For use with (application)
- Ion concentration titrations including salt in food products, chloride in drinking water and wastewater.
- Includes
- 9780SC silver billet electrode, 91CBNC electrode cable, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, literature on USB drive, 110-240V power supplies
- Inputs
- BNC (ion selective electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), stirrer input (stirrer probe)
- IP rating
- IP51
- Languages
- Chinese, English, French, German, Italian, Japanese, Korean, Portuguese and Spanish
- Material
- ABS Plastic
- Memory
- Non-volatile memory preserves data logs and settings
- Methods
- Up to 10 methods with optional password protection; import/export methods via USB flash drive
- Model
- Orion Star T930
- Outputs
- 3 USB ports for printer, USB flash drive and computer software connections
- Precision
- 0.5% RSD, dependent on environment and handling conditions
- Preset endpoints
- 1 to 3
- Probe type
- Ion selective electrode (ISE), reference half-cell electrode, ATC temperature probe, stirrer probe
- Range (ISE)
- 0.0001 to 19990
- Range (mV)
- -2000.0 to +2000.0 mV
- Relative humidity range
- 5 to 85%, non-condensing
- Resolution (ISE)
- 0.0001 minimum, 1 to 4 significant digits (user selectable)
- Resolution (mV)
- 0.1 mV
- Resolution (temperature)
- 0.1 °C
- Sample ID
- Auto-incremental, manual or off
- Stirrer probe speeds
- 5 user selectable speeds (250 to 3700 RPM)
- Temperature (english) operating
- 41°F to 104°F
- Temperature (metric) operating
- 5°C to 40°C
- Temperature probe calibration
- Optional 1 point offset calibration
- Temperature range
- -5.0°C to +100.0°C
- Temperature selection
- Manual or automatic
- Time and date
- Yes, with nonvolatile battery backup
- Titrant determination
- Standardization titration or manual concentration input
- Titration process control
- Routine, quick, careful or user defined
- Titration setup wizard
- Yes, using Start a New Titration workflow
- Titration technique
- Equivalence point or preset endpoint
- Titration types
- Direct or back titration
- Updateable firmware
- Yes, using USB flash drive
- Warranty
- 1 year replacement warranty
- Dimensions (L x W x H)
- 10 x 16 x 14 in. (25.4 x 40.6 x 35.6 cm)
- Electrical requirements
- 100-240V, 50/60 Hz
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- START9300Orion Star T930 ion titrator without electrode
- START9102Orion Star T910 pH titrator ROSS Sure-Flow kit, 8172BNWP ROSS Sure-Flow pH electrode, 927007MD ATC probe, 20 mL burette, stirrer probe, dispenser probe
- START9100Orion Star T910 pH titrator without electrode, 20 mL burette, stirrer probe, dispenser probe, standard tubing kit, 1 L plastic bottle
- START9101Orion Star T910 pH titrator standard ROSS kit, 8102BNUWP ROSS Ultra pH electrode, 927007MD ATC probe, 20 mL burette, stirrer probe, dispenser probe
- START9201Orion Star T920 redox titrator standard kit, 9778BNWP glass-body redox electrode, 20mL burette, stirrer probe, dispenser probe, standard tubing kit
- START9200Orion Star T920 redox titrator without electrode, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1 L plastic bottle
- START9402Orion Star T940 all-in-one titrator ROSS Sure-Flow kit, 8172BNWP ROSS Sure-Flow pH electrode, 927007MD ATC probe, 20 mL burette, stirrer probe
- START9400Orion Star T940 all-in-one titrator without electrode, 20 mL burette, stirrer probe, dispenser probe, standard tubing kit, 1 L plastic bottle
- START9401Orion Star T940 all-in-one titrator standard ROSS kit, 8102BNUWP ROSS Ultra pH electrode, 927007MD ATC probe, 20 mL burette, stirrer probe