BK-AA220AES
Atomic Absorption Spectrophotometer, power supply AC220 V, 50 Hz
- Manufacturer
- BIOBASE
- Reference
- BK-AA220AES
- Product line
- Atomic Absorption Spectrophotometer
- Item type
- UV-Vis spectrophotometer (cuvette)
Specifications
- Wavelength Range
- 170 to 900 nm
- Spectral Bandwidth
- 0.1, 0.2, 0.4, 1.6 nm
- Wavelength Accuracy
- ≤±0.1 nm
- Wavelength Repeatability
- ≤0.05 nm
- Detector
- Photomultiplier tube detector
- Baseline Stability
- Drift ≤0.002Abs, noise ≤0.001Abs(Steady 30min) Drift ≤0.002Abs, noise ≤0.001Abs(Dynamic 15min)
- Resolution
- Spectral bandwidth deviation ≤0.02 nm Manganese double line valley to peak energy ratio ≤25%
- Characteristic Concentration/Content
- Flame method for Cu: ≤0.025μg/mL/1% Graphite furnace method for Cd: ≤0.5pg
- Detection Limit
- Flame method for Cu: ≤0.002μg/mL Graphite furnace method for Cd: ≤0.4pg
- Preciseness
- Flame method for Cu: ≤0.3% Graphite furnace method for Cd: ≤2.0%
- Background Correction Ability
- Deuterium lamp calibration 1.0Abs, background ≥90 times Self-absorption method correction 2.0Abs, background ≥90 times
- Diffraction Grating
- Groove 1800lines/mm, blaze wavelength 250 nm
- Lamp Stand
- 8(Support simultaneous preheating of 1 to 4 lights)
- Standard Accessory
- Element lamp(Cu*1, Cd*1, Hg*1), software working station
- Optional Accessory
- Wide slit combustion head, adjustable nebulizer, hydride vapor generator, audit trailing software, oil-free air compressor, recirculating cooling water system, computer, printer, autosampler, oxygen-enriched high-temperature flame
- Power Supply
- AC220 V, 50 Hz
- External Size(W*D*H)
- Main unit: 1020 x 530 x 500 mm GF power supply: 350 x 530 x 430 mm
- Package Size(W*D*H)
- Main unit: 1280 x 750 x 830 mm GF power supply: 610 x 670 x 810 mm
- Net Weight
- Main unit: 85 kg GF power supply: 32 kg
- Gross Weight
- Main unit: 160 kg GF power supply: 70 kg
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- BK-AA180BAtomic Absorption Spectrophotometer, power supply AC220 V, 50/60 Hz; 110 V, 60 Hz(Standard)
- BK-AA4530FAtomic Absorption Spectrophotometer, power supply 220 V, 50 Hz
- BK-AA3618Atomic Absorption Spectrophotometer, power supply AC 220 V, 50/60 Hz
- BK-AA3638Atomic Absorption Spectrophotometer, power supply AC 220 V, 50/60 Hz
- BK-AA3668Atomic Absorption Spectrophotometer, power supply AC 220 V, 50/60 Hz
- BK-AA320NAtomic Absorption Spectrophotometer, power supply 220 V, 50 Hz