BK-AA3638
Atomic Absorption Spectrophotometer, power supply AC 220 V, 50/60 Hz
- Manufacturer
- BIOBASE
- Reference
- BK-AA3638
- Product line
- Atomic Absorption Spectrophotometer
- Item type
- UV-Vis spectrophotometer (cuvette)
Specifications
- Instrument Type
- Flame & GFAA integrated system
- Optical System Beam Type
- Double-beam
- Optical System Diffraction Grating
- Groove 1800lines/mm, blaze wavelength 250 nm
- Optical System Wavelength Range
- 185 to 900 nm
- Optical System Spectral Bandwidth
- 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 nm
- Optical System Wavelength Accuracy
- ±0.2 nm
- Optical System Wavelength Repeatability
- ≤0.1 nm
- Optical System Spectral Bandwidth Deviation
- ±0.02 nm
- Optical System Lamp Stand
- 8(Support simultaneous preheating of 1 to 4 lights)
- Photometric Performance Reading Modes
- Transmittance, absorbance, concentration
- Photometric Performance Photometric Range
- 0 to 125%T, -0.1 to 3.00Abs
- Photometric Performance Baseline Stability
- Drift ±0.003Abs(Steady 30min) Drift ±0.005Abs(Dynamic 15min)
- Photometric Performance Background Correction Ability
- Deuterium lamp: ≥40 fold(at 1Abs)
- Flame Analysis Characteristic Concentration
- Flame method for Cu: ≤0.035μg/mL
- Flame Analysis Detection Limit
- Flame method for Cu: ≤0.006μg/mL
- Flame Analysis Repeatability
- Flame method for Cu: ≤0.6%
- Flame Analysis Burner
- Titanium burner head
- Flame Analysis Nebulizer
- High-efficiency glass nebulizer
- Flame Analysis Safety Features
- Flame gas & auxiliary gas abnormal pressure protection
- Graphite Furnace Analysis Characteristic Mass
- Cd ≤0.6pg Cu ≤20pg
- Graphite Furnace Analysis Detection Limit
- Cd ≤1.0pg Cu ≤25pg
- Graphite Furnace Analysis Repeatability(Cd)
- ≤3%
- Graphite Furnace Analysis Graphite Furnace Temperature Range
- RT to 3000 °C
- Graphite Furnace Analysis Heating Rate
- Maximum 2000 °C/s
- Graphite Furnace Analysis Safety Features
- Alarms for cooling water flow, protective gas pressure, furnace body temperature, power supply temperature, and graphite tube installation
- Data Processing Calculation Method for Concentration
- Standard curve method(6 linear and non-linear fitting methods), standard addition method, interpolation method
- Data Processing Number of Repeated Measurements
- 1 to 30 times; calculates and provides the average value, standard deviation and relative standard deviation of absorbance and concentration
- Standard Accessory
- Element lamp(Cu*1, Cd*1), oil-free air compressor, software, U disk, graphite tube*5, pipette*1(5 to 50μl), pipette tip*1000(200μl)
- Optional Accessory
- Oxygen-enriched high-temperature flame, hydride generator, audit trailing software, recirculating chiller, graphite furnace autosampler, flame autosampler
- Power Supply
- AC 220 V, 50/60 Hz
- External Size(W*D*H)
- 830 x 650 x 560 mm
- Package Size(W*D*H)
- 1010 x 780 x 830 mm
- Net Weight
- 120 kg
- Gross Weight
- 183 kg
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- BK-AA180BAtomic Absorption Spectrophotometer, power supply AC220 V, 50/60 Hz; 110 V, 60 Hz(Standard)
- BK-AA4530FAtomic Absorption Spectrophotometer, power supply 220 V, 50 Hz
- BK-AA3618Atomic Absorption Spectrophotometer, power supply AC 220 V, 50/60 Hz
- BK-AA3668Atomic Absorption Spectrophotometer, power supply AC 220 V, 50/60 Hz
- BK-AA220AESAtomic Absorption Spectrophotometer, power supply AC220 V, 50 Hz
- BK-AA320NAtomic Absorption Spectrophotometer, power supply 220 V, 50 Hz