BK-AA3668

Atomic Absorption Spectrophotometer, power supply AC 220 V, 50/60 Hz

Manufacturer
BIOBASE
Reference
BK-AA3668

Specifications

Instrument Type
Flame & GFAA integrated system
Optical System Beam Type
Double-beam
Optical System Diffraction Grating
Groove 1800lines/mm, blaze wavelength 250 nm
Optical System Wavelength Range
185 to 900 nm
Optical System Spectral Bandwidth
0.1, 0.2, 0.4, 0.7, 1.4, 2.0 nm
Optical System Wavelength Accuracy
±0.15 nm
Optical System Wavelength Repeatability
≤0.05 nm
Optical System Spectral Bandwidth Deviation
±0.02 nm
Optical System Lamp Stand
8(Support simultaneous preheating of 1 to 4 lights)
Photometric Performance Reading Modes
Transmittance, absorbance, concentration
Photometric Performance Photometric Range
0 to 125%T, -0.1 to 3.00Abs
Photometric Performance Baseline Stability
Drift ±0.002Abs(Steady 30min) Drift ±0.004Abs(Dynamic 15min)
Photometric Performance Background Correction Ability
Deuterium lamp: ≥40 fold(at 1Abs) Self-reversal: ≥80 fold(at 1Abs)
Flame Analysis Characteristic Concentration
Flame method for Cu: ≤0.02μg/mL
Flame Analysis Detection Limit
Flame method for Cu: ≤0.004μg/mL
Flame Analysis Repeatability
Flame method for Cu: ≤0.5%
Flame Analysis Burner
Titanium burner head
Flame Analysis Nebulizer
High-efficiency glass nebulizer
Flame Analysis Safety Features
Flame gas & auxiliary gas abnormal pressure protection
Graphite Furnace Analysis Characteristic Mass
Cd ≤0.5pg Cu ≤20pg
Graphite Furnace Analysis Detection Limit
Cd ≤0.8pg Cu ≤25pg
Graphite Furnace Analysis Repeatability(Cd)
≤2.5%
Graphite Furnace Analysis Graphite Furnace Temperature Range
RT to 3000 °C
Graphite Furnace Analysis Heating Rate
Maximum 2000 °C/s
Graphite Furnace Analysis Safety Features
Alarms for cooling water flow, protective gas pressure, furnace body temperature, power supply temperature, and graphite tube installation
Data Processing Calculation Method for Concentration
Standard curve method(6 linear and non-linear fitting methods), standard addition method, interpolation method
Data Processing Number of Repeated Measurements
1 to 30 times; calculates and provides the average value, standard deviation and relative standard deviation of absorbance and concentration
Standard Accessory
Element lamp(Cu*1, Cd*1), oil-free air compressor, software, U disk, graphite tube*5, pipette*1(5 to 50μl), pipette tip*1000(200μl)
Optional Accessory
Oxygen-enriched high-temperature flame, hydride generator, audit trailing software, recirculating chiller, graphite furnace autosampler, flame autosampler
Power Supply
AC 220 V, 50/60 Hz
External Size(W*D*H)
830 x 650 x 560 mm
Package Size(W*D*H)
1010 x 780 x 830 mm
Net Weight
125 kg
Gross Weight
183 kg

Add this reference to my quote