Product
Ludlum Model 23, mrem electronic personal dosimeter
Ludlum part number
51-2958
Radiation Detected
Gamma and X-ray (35 keV to 3 MeV)
Detector
Silicon semiconductor
Measurement and Display Range
Accumulated Dose
Dose Rate*
0.1 mrem/hr to 99.99 rem/hr
Display
4-digit OLED with indicators for dose, dose rate, operating time, and alarm
Accuracy
≤ ±10% (0.01 to 999.9 mSv (1.0 mrem to 99.99 rem), 137 Cs)
Energy Response
≤ ±20% (60 keV to 1.5 MeV, 137 Cs)
Angular Response
≤ ±20% (Up to ± 60 degree vertical and horizontal, 137 Cs)
Dose Rate Linearity*
≤ ±10% ( 137 Cs)
Alarm Volume
Approximately 60 dB (at 20 cm)
Data Logging
Up to 600 records (optional IR reader required for data transmission to PC)
Environment
-10 to 40 °C (14 to 104 °F) ≤ 90% relative humidity (non-condensing)
Power
One each coin-type lithium battery (CR2450, P/N 21-8639)
Battery Life
720 hours (without alarm / communication)
Size (H x W x D)
112.8 x 30 x 12 mm (4.4 x 1.2 x 0.5 in.)
Weight
55.9 g (2 oz.)