Model AT-725
Ludlum Model AT-725, pencil dosimeter
- Manufacturer
- Ludlum Measurements
- Reference
- Model AT-725
- Product line
- Personal dosimeters and radiation pagers
Specifications
- Product
- Ludlum Model AT-725, pencil dosimeter
- Ludlum part number
- 51-2939
- Range
- 0 to 5 R
- Energy Response
- 16 keV to 2 MeV
- Radiation Detected
- Gamma and X-ray from 16 keV to 6 MeV
- Detector
- Fiber electrometer mounted in an electrically conductive plastic ion chamber
- Detector Housing
- Very low permeability plastic, hermetically sealed
- Accuracy
- Within 10% of true exposure
- Rate Response
- Dose rate independent for gamma and X-ray radiation
- Electrical Leakage
- Less than 1.0% of full scale for 24 hours at 50 °C (122 °F)
- Temperature Range
- -20 to 50 °C (-4 to 122 °F)
- Relative Humidity
- Up to 90%
- Finish
- Barrel and End Caps
- Clip
- Color coded plastic (color signifies range) or metal clip
- Size (D x L)
- 1.5 x 12.4 cm (0.6 x 4.5 in.)
- Weight
- 25 g (1.0 oz)
- Warranty
- 2 year limited warranty
- Qualifications
- Meets ANSI Specifications N13.5 and N322-1997
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 23Ludlum Model 23, mrem electronic personal dosimeter
- Model 23-1Ludlum Model 23-1, mSv electronic personal dosimeter
- Model AT-138Ludlum Model AT-138, pencil dosimeter
- Model AT-138SLudlum Model AT-138S, pencil dosimeter
- Model 25-ISLudlum Model 25-IS, internal energy-compensated gm
- Model 25Ludlum Model 25, internal energy-compensated gm