Model 25-IS
Ludlum Model 25-IS, internal energy-compensated gm
- Manufacturer
- Ludlum Measurements
- Reference
- Model 25-IS
- Product line
- Personal dosimeters and radiation pagers
Specifications
- Product
- Ludlum Model 25-IS, internal energy-compensated gm
- Ludlum part number
- m25-is-series
- Detector
- Internal energy-compensated GM
- Energy Range
- 60 keV to 2 MeV
- Display
- 3½ digit backlit LCD display
- Display Range
- Model 25-IS: 0.01 mR/hr (mR) to 999 R/hr (R) Model 25-IS-1: 0.001 mSv/h (mSv) to 9.99 Sv/h (Sv) Also displays time remaining from 19:59 to 00:01 (in hh:mm format)
- Gamma Response
- Model 25-IS: Typically 18 cpm per mR/hr Model 25-IS-1: Typically 1800 cpm per mSv/hr
- Alerts & Alarms
- Radiation alarms are adjustable over entire range. There are distinct alerts and alarms for exposure and accumulated dose: Dose rate alert Dose rate (hi) Accumulated dose alert Accumulated dose (hi) Time remaining to all
- Loss of Count Protection
- If no pulses are received from the detector in 60 minutes, instrument failure is indicated by an audible and visual alarm and the display alternating between "0.0" and "F"
- Audio
- Built-in speaker (typically 95 dB at 30.5 cm [1 ft])
- Calibration
- Requires no tools or software when exposed to a traceable radiation field
- Calibrated Range
- Normally calibrated from: Model 25-IS: 2 mR/hr to 999 R/hr Model 25-IS-1: 0.02 mSv/h to 9.99 Sv/h
- Temperature Range
- -40 to 65 °C (-40 to 150 °F)
- Power
- 2 each lithium coin cell batteries (see Warning note below)
- Battery Life
- Typically 6000 hours. Provides 8 hours warning of low battery.
- Construction
- Injection-molded plastic housing with sub-surface printed membrane front panel, completely gasketed for water resistance
- Environmental Rating
- IP67
- Size (H x W x L)
- 7.6 x 5.4 x 1.7 cm (3.0 x 2.1 x 0.69 in.)
- Weight
- 144 g (5.1 oz), including batteries
- Intrinsically Safe Certification
- Class I, Division 1, Groups A-D, Class II, III, Division 1, Groups E, F, and G, T6 UL 61010-1/CSA C22.2 No. 61010-1: Standard for Safety Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1: General
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 23Ludlum Model 23, mrem electronic personal dosimeter
- Model 23-1Ludlum Model 23-1, mSv electronic personal dosimeter
- Model AT-138Ludlum Model AT-138, pencil dosimeter
- Model AT-138SLudlum Model AT-138S, pencil dosimeter
- Model AT-725Ludlum Model AT-725, pencil dosimeter
- Model 25Ludlum Model 25, internal energy-compensated gm