Model 44-10

Ludlum Model 44-10, gamma detector

Manufacturer
Ludlum Measurements
Reference
Model 44-10

Specifications

Product
Ludlum Model 44-10, gamma detector
Ludlum part number
47-1540
Indicated Use
Low-level, wide-energy gamma detection
Detector Type
NaI(Tl) scintillator, 5.1 cm (2 in.) dia. x 5.1 cm (2 in.) thick
Efficiency (4π)
125 I: 4% 57 Co: 20% 137 Cs: 9% 60 Co: 15%
Typical Sensitivity ( 137 Cs gamma)
900 cpm per μR/hr
Recommended Energy Range
30 keV to 3.0 MeV
Energy Response
Energy dependent (see Energy Response Curve )
Typical Background (10 µR/hr field)
9750 cpm
Photomultiplier Tube
5.1 cm (2 in.) diameter
Operating Voltage
500 to 1200 V
Temperature Range
-20 to 50 °C (-4 to 122 °F) May be certified to operate from -40 to 65 °C (-40 to 150 °F)
Environmental Rating
IP65
Connector
Series "C"
Construction
Aluminum with beige powder coat
Size (D x L)
6.6 x 27.9 cm (2.6 x 11 in.)
Weight
1.04 kg (2.3 lb)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line