Model 44-11
Ludlum Model 44-11, gamma detector
- Manufacturer
- Ludlum Measurements
- Reference
- Model 44-11
- Product line
- Detector probes: alpha, beta and gamma
Specifications
- Product
- Ludlum Model 44-11, gamma detector
- Ludlum part number
- 47-1541
- Indicated Use
- High sensitivity gamma measurements, Gamma survey
- Detector Type
- NaI(Tl) scintillator, 5.1 cm (2 in) dia. x 5.1 cm (2 in) thick
- Typical Sensitivity ( 137 Cs gamma)
- 900 cpm per µR/hr
- Energy Response
- Energy dependent
- Typical Background (10 µR/hr field)
- 9750 cpm
- Photomultiplier Tube
- 5.1 cm (2 in) diameter, magnetically shielded
- Operating Voltage
- 500 to 1200 V (typical)
- Connector
- Series "C"
- Construction
- Integral-line
- Housing Construction
- 0.05 cm (0.02 in.) thick aluminum
- Size (D x L)
- 6.4 x 26.7 cm (2.5 x 10.5 in)
- Weight
- 0.6 kg (1.4 lb)
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 44-10Ludlum Model 44-10, gamma detector
- Model 44-132Ludlum Model 44-132, gamma scintillator detector
- Model 44-159-1Ludlum Model 44-159-1, CsI scintillator detector
- Model 44-2Ludlum Model 44-2, gamma detector
- Model 44-20Ludlum Model 44-20, gamma detector
- Model 44-62Ludlum Model 44-62, gamma detector
- Model 43-1-1Ludlum Model 43-1-1, alpha-beta detector
- Model 43-135-1Ludlum Model 43-135-1, large area 2π proportional detector
- Model 43-143-1Ludlum Model 43-143-1, alpha-beta detector
- Model 43-147Ludlum Model 43-147, alpha-beta detector
- Model 43-2-2Ludlum Model 43-2-2, alpha-beta detector
- Model 43-37Ludlum Model 43-37