Model 44-2

Ludlum Model 44-2, gamma detector

Manufacturer
Ludlum Measurements
Reference
Model 44-2

Specifications

Product
Ludlum Model 44-2, gamma detector
Ludlum part number
47-1532
Indicated Use
Low-level, wide-energy gamma detection
Detector Type
NaI(Tl) scintillator, 2.5 cm (1 in) dia. x 2.5 cm (1 in) thick
Efficiency (4π)
125 I: 7% 57 Co: 10% 137 Cs: 3% 60 Co: 3%
Typical Sensitivity ( 137 Cs gamma)
175 cpm per μR/hr
Recommended Energy Range
20 keV to 1.5 MeV
Exposure Energy Response (gamma)
Energy dependent (see Response Curve )
H*(10) Ambient Dose Equivalent Energy Response
Energy dependent (see Response Curve)
Typical Background (10 µR/hr field)
1800 cpm
Photomultiplier Tube
2.86 cm (1.123 in) diameter, magnetically shielded
Operating Voltage
500 to 1200 V
Temperature Range
-15 to 50 °C (5 to 122 °F) May be certified for -40 to 65 °C (-40 to 150 °F)
Environmental Rating
IP65
Connector
Series "C"
Construction
Aluminum with beige powder coat
Size (D x L)
5.1 x 21 cm (2 x 8.25 in)
Weight
0.5 kg (1 lb)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item