Model 133
Ludlum Model 133, gamma area monitoring
- Manufacturer
- Ludlum Measurements
- Reference
- Model 133
- Product line
- Detector probes: alpha, beta and gamma
Specifications
- Product
- Ludlum Model 133, gamma area monitoring
- Ludlum part number
- 133-series_2
- Part Number
- 133-2: 47-1717 133-2-1: 47-2450 †
- Indicated Use
- Gamma area monitoring
- Detector Type
- Energy compensated, halogen quenched GM, 30 mg/cm 2 stainless steel
- Linear Range
- Without DTC*: 1 µSv/h to 2 mSv/h (0.1 to 200 mR/hr) With DTC*: 1 µSv/h to 10 mSv/h (0.1 mR/hr to 1 R/hr) With SCC or DTC2**: 1 µSv/h to 50 mSv/h (0.1 mR/hr to 5 R/hr)
- Typical Sensitivity ( 137 Cs gamma)
- 100 cpm per µSv/h (1000 cpm per mR/hr)
- Energy Response (60 keV - 3 MeV)
- Within 25% of true value (see Response Curves below)
- Typical Background (10 µR/hr field)
- 12 cpm
- Typical Dead Time
- 50 µs
- Operating Voltage
- 550 Vdc
- Connector
- 133-2: Series "C". Other connectors are available upon request. 133-2-1: 30 m (100 ft) hard-wired cable
- Size (Dia x L)
- 2.5 x 10.2 cm (1 x 4 in.)
- Weight
- 0.090 kg (0.20 lb)
- Part Number
- 133-3: 47-1314 133-3-1: 47-4387 †
- Indicated Use
- Gamma area monitoring
- Detector Type
- Energy compensated, halogen quenched GM, 30 mg/cm 2 stainless steel
- Linear Range
- Without DTC*: 1 µSv/h to 10 mSv/h (0.1 mR/hr to 1 R/hr) With DTC*: 1 µSv/h to 50 mSv/h (0.1 mR/hr to 5 R/hr) With SCC or DTC2**: 1 µSv/h to 250 mSv/h (0.1 mR/hr to 25 R/hr)
- Typical Sensitivity ( 137 Cs gamma)
- 45 cpm per µSv/h (450 cpm per mR/hr)
- Energy Response (60 keV - 3 MeV)
- Within 25% of true value
- Typical Background (10 µR/hr field)
- 6 cpm
- Typical Dead Time
- 50 µs
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 44-10Ludlum Model 44-10, gamma detector
- Model 44-11Ludlum Model 44-11, gamma detector
- Model 44-132Ludlum Model 44-132, gamma scintillator detector
- Model 44-159-1Ludlum Model 44-159-1, CsI scintillator detector
- Model 44-2Ludlum Model 44-2, gamma detector
- Model 44-20Ludlum Model 44-20, gamma detector
- Model 44-62Ludlum Model 44-62, gamma detector
- Model 43-1-1Ludlum Model 43-1-1, alpha-beta detector
- Model 43-135-1Ludlum Model 43-135-1, large area 2π proportional detector
- Model 43-143-1Ludlum Model 43-143-1, alpha-beta detector
- Model 43-147Ludlum Model 43-147, alpha-beta detector
- Model 43-2-2Ludlum Model 43-2-2, alpha-beta detector