Model 43-68
Ludlum Model 43-68, alpha-beta detector
- Manufacturer
- Ludlum Measurements
- Reference
- Model 43-68
- Product line
- Detector probes: alpha, beta and gamma
Specifications
- Product
- Ludlum Model 43-68, alpha-beta detector
- Ludlum part number
- 47-2005
- Indicated Use
- Alpha-beta survey
- Detector Type
- Rechargable gas proportional
- Recommended Counting Gas
- P-10 (10% methane, 90% argon)
- Gas Recharge
- Will operate on static charge for at least 15 hours with 1 m (39 in.) cable; 5-hour static charge if connected to a simultaneous alpha/beta measuring instrument
- Window
- 0.8 mg/cm 2 metalized polyester (other thicknesses available)
- Window Area
- Active: 126 cm 2 (19.5 in 2 ) Open: 100 cm 2 (15.6 in 2 )
- Efficiency (4π)
- 239 Pu: 20% 14 C: 15% 99 Tc: 23% 90 Sr/ 90 Y: 20% Gamma: 1%
- Simultaneous Alpha-Beta Counting
- 239 Pu: 17.5% 99 Tc: 16% 90 Sr/ 90 Y: 20%
- Typical Background (10 µR/hr field)
- Alpha: ≤ 3 cpm when operating at alpha-only plateau region Beta-Gamma: 350 cpm
- Operating Voltage
- Alpha: 1100 to 1400 V (typical) Beta-Gamma: 1600 to 1800 V (typical)
- Counter Threshold Setting
- 2 to 5 mV (typical)
- Gas Connectors
- double-end quick disconnect
- Size (H x W x L)
- 10 x 11.7 x 19.8 cm (3.9 x 4.6 x 7.8 in.) with handle
- Weight
- 0.9 kg (2 lb)
- Cables
- The maximum recommended cable lengths for alpha-beta gas proportional detectors is 3.66 m (12 ft). Longer lengths can be used, but detector performance will be reduced.
- Model 3002
- Digital Survey Meter For alpha-beta counting Stores up to 1000 data points
- Model 12
- Analog Survey Meter
- Model 16
- Analog Survey Meter Adjustable window
- Model 18
- Analog Survey Meter Adjustable window 3 detector setups
- Model 2000
- Scaler Scaler Range: 0 to 999,999 counts Timer: 0 to 999 minutes
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 44-10Ludlum Model 44-10, gamma detector
- Model 44-11Ludlum Model 44-11, gamma detector
- Model 44-132Ludlum Model 44-132, gamma scintillator detector
- Model 44-159-1Ludlum Model 44-159-1, CsI scintillator detector
- Model 44-2Ludlum Model 44-2, gamma detector
- Model 44-20Ludlum Model 44-20, gamma detector
- Model 44-62Ludlum Model 44-62, gamma detector
- Model 43-1-1Ludlum Model 43-1-1, alpha-beta detector
- Model 43-135-1Ludlum Model 43-135-1, large area 2π proportional detector
- Model 43-143-1Ludlum Model 43-143-1, alpha-beta detector
- Model 43-147Ludlum Model 43-147, alpha-beta detector
- Model 43-2-2Ludlum Model 43-2-2, alpha-beta detector