Model 43-89

Ludlum Model 43-89, alpha-beta detector

Manufacturer
Ludlum Measurements
Reference
Model 43-89

Specifications

Product
Ludlum Model 43-89, alpha-beta detector
Ludlum part number
47-2430
Indicated Use
Simultaneous alpha and beta survey
Detector Type
ZnS(Ag) adhered to 0.25 mm (0.01 in) thick plastic scintillator
Window
1.2 mg/cm 2 metalized polyester
Window Area
Active: approx. 125 cm 2 (19.4 in 2 ) Open: approx. 100 cm 2 (15.5 in 2 )
Protective Screen
0.76 mm (0.030 in.) thick stainless steel, 6.4 mm (0.25 in.) hex, 79% open Outer Dimensions (L x W): 168 x 92 mm (6.6 x 3.6 in.)
Efficiency (4π)
239 Pu: 17.5% 99 Tc: 10% 90 Sr/ 90 Y: 17.5%
Typical Background (10 µR/hr field)
Alpha: 3 cpm or less Beta: 300 cpm or less
Crosstalk
Alpha to Beta: 10% or less Beta to Alpha: 1% or less
Detector Uniformity
Within 10% of average reading
Gamma Sensitivity
approx. 20 cpm per µR/hr
Photomultiplier Tube
38 mm (1.5 in.) diameter
Operating Voltage
Typically 500 to 1000 V (1500 V max.)
Temperature Range
-20 to 50 °C (-4 to 122 °F)
Environmental Rating
IP rating of 63
Connector
Series "C" (others available)
Construction
Aluminum with beige powder coating
Size (H x W x L)
10 x 10 x 33 cm (4 x 4 x 12.25 in.)
Weight
0.68 kg (1.5 lbs)
Model 3002
Digital Survey Meter For alpha-beta counting Can store up to 1000 data points
Model 3003 Series
Multi-Detector Digital Survey Meter Up to 4 selectable & configurable detector settings Alpha, beta, or alpha+beta measurements Single channel analyzer Can store up to 1000 data points

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.