Model 43-93

Ludlum Model 43-93, alpha-beta detector

Manufacturer
Ludlum Measurements
Reference
Model 43-93

Specifications

Product
Ludlum Model 43-93, alpha-beta detector
Ludlum part number
47-2556
Indicated Use
Simultaneous alpha and beta survey
Detector Type
ZnS(Ag) scintillator adhered to 0.25 mm (0.01 in) thick plastic scintillator
Window
1.2 mg/cm 2 metalized polyester
Window Area
Active: 100 cm 2 (15.5 in 2 ) Open: 88 cm 2 (13.6 in 2 )
Protective Screen
0.79 mm (0.031 in.) thick, 6.5 mm (0.26 in.) square openings, 88% open Outer Dimensions (L x W): 168 x 92 mm (6.6 x 3.6 in.) Tool-less removal for quick window replacements
Efficiency (4π)
99 Tc: 15% 90 Sr/ 90 Y: 20% 239 Pu: 20%
Typical Background (10 µR/hr field)
Alpha: 3 cpm or less Beta: 300 cpm or less
Cross Talk
Alpha to Beta: 10% or less Beta to Alpha: 1% or less
Detector Uniformity
Within 15% of average reading
Gamma Sensitivity
15 to 20 cpm per µR/hr ( 137 Cs)
Photomultiplier Tube
29 mm (1.125 in.) diameter
Operating Voltage
Typically 500 to 1200 V (1500 V max.)
Temperature Range
-20 to 40 °C (-4 to 104 °F) A version designed to be used at temperatures up to 50 °C (122 °F) is also available. See Options .
Connector
Series "C" (others available)
Construction
Aluminum with beige powder coating
Size (H x W x L)
6.4 x 9.5 x 31 cm (2.5 x 3.8 x 12.3 in.)
Weight
0.45 kg (1.0 lb)
Model 3002
Digital Survey Meter For alpha-beta counting Can store up to 1000 data points
Model 3003 Series
Multi-Detector Digital Survey Meter Up to 4 selectable & configurable detector settings Alpha, beta, or alpha+beta measurements Single channel analyzer Can store up to 1000 data points
Model 2360
Analog Datalogging Meter with Digital Scaler LCD Simultaneous alpha-beta measurements Can store up to 550 data points

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item

Other references in this line