Model 44-183
Ludlum Model 44-183, shielded gamma detector
- Manufacturer
- Ludlum Measurements
- Reference
- Model 44-183
- Product line
- Detector probes: alpha, beta and gamma
Specifications
- Product
- Ludlum Model 44-183, shielded gamma detector
- Ludlum part number
- 47-3758
- Indicated Use
- Gamma survey
- Detector Type
- Energy-compensated GM with lead shielding
- Linear Range
- Without DTC*: 10 μSv/h to 20 mSv/h (1 mR/hr to 2 R/hr) With DTC*: 10 μSv/h to 100 mSv/h (1 mR/hr to 10 R/hr)
- Typical Sensitivity ( 137 Cs)
- 0.17 cps per μSv/h (100 cpm per mR/hr)
- Energy Response (60 keV - 3 MeV)
- Within 15% of true value
- Front to Back Shield Ratio
- 10 : 1
- Typical Dead Time
- 50 µs
- Operating Voltage
- 550 V (reccomended)
- Input Sensitivity
- -30 mV ±10 mV
- Connector
- Series "C"
- Construction
- Aluminum and lead with painted exterior
- Temperature Range
- -20 to 50 °C (-4 to 122 °F)
- Size (H x W x L)
- 10.7 x 12.2 x 12.7 cm (4.2 x 4.8 x 5.0 in.)
- Weight
- 3.3 kg (7.3 lb)
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 44-10Ludlum Model 44-10, gamma detector
- Model 44-11Ludlum Model 44-11, gamma detector
- Model 44-132Ludlum Model 44-132, gamma scintillator detector
- Model 44-159-1Ludlum Model 44-159-1, CsI scintillator detector
- Model 44-2Ludlum Model 44-2, gamma detector
- Model 44-20Ludlum Model 44-20, gamma detector
- Model 44-62Ludlum Model 44-62, gamma detector
- Model 43-1-1Ludlum Model 43-1-1, alpha-beta detector
- Model 43-135-1Ludlum Model 43-135-1, large area 2π proportional detector
- Model 43-143-1Ludlum Model 43-143-1, alpha-beta detector
- Model 43-147Ludlum Model 43-147, alpha-beta detector
- Model 43-2-2Ludlum Model 43-2-2, alpha-beta detector