Model 44-26
Ludlum Model 44-26, alpha-beta-gamma foot monitor detector
- Manufacturer
- Ludlum Measurements
- Reference
- Model 44-26
- Product line
- Detector probes: alpha, beta and gamma
Specifications
- Product
- Ludlum Model 44-26, alpha-beta-gamma foot monitor detector
- Ludlum part number
- 47-1509
- Indicated Use
- Alpha, beta, gamma foot monitor
- Detector Type
- 3 pancake type, halogen quenched GMs
- Window
- 1.7 ± 0.3 mg/cm 2 mica
- Window Area
- Active: 46 cm 2 (7.1 in 2 ) Open: 27 cm 2 (4.2 in 2 )
- Efficiency (4π)
- 14 C: 5% 90 Sr/ 90 Y: 22% 99 Tc: 19% 32 P: 32% 239 Pu: 15%
- Energy Response
- Energy dependent
- Typical Sensitivity
- 9900 cpm per mR/hr
- Typical Background (10 µR/hr field)
- 180 cpm
- Operating Voltage
- 900 V
- Construction
- Aluminum housing
- Size (H x W x L)
- 27.4 x 17.8 x 22.9 cm (10.8 x 7 x 9 in.)
- Weight
- 2 kg (4.4 lb)
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 44-10Ludlum Model 44-10, gamma detector
- Model 44-11Ludlum Model 44-11, gamma detector
- Model 44-132Ludlum Model 44-132, gamma scintillator detector
- Model 44-159-1Ludlum Model 44-159-1, CsI scintillator detector
- Model 44-2Ludlum Model 44-2, gamma detector
- Model 44-20Ludlum Model 44-20, gamma detector
- Model 44-62Ludlum Model 44-62, gamma detector
- Model 43-1-1Ludlum Model 43-1-1, alpha-beta detector
- Model 43-135-1Ludlum Model 43-135-1, large area 2π proportional detector
- Model 43-143-1Ludlum Model 43-143-1, alpha-beta detector
- Model 43-147Ludlum Model 43-147, alpha-beta detector
- Model 43-2-2Ludlum Model 43-2-2, alpha-beta detector