Model 44-3

Ludlum Model 44-3, gamma detector

Manufacturer
Ludlum Measurements
Reference
Model 44-3

Specifications

Product
Ludlum Model 44-3, gamma detector
Ludlum part number
47-1533
Indicated Use
125 I and low-energy gamma survey
Detector Type
NaI(Tl) scintillator, 2.5 cm (1 in) dia. x 0.1 cm (0.04 in) thick
Window
18.4 mg/cm 2
Window Area
Active: 5 cm 2 (0.8 in 2 ) Open: 5 cm 2 (0.8 in 2 )
Efficiency (4π)
125 I: 33.5% (derived from 129 I efficiency) 129 I: 18%
Typical Sensitivity ( 125 I)
675 cpm per µR/hr
Typical Background (10 µR/hr field)
300 cpm
Recommended Energy Range
10 to 60 keV
Energy Response
Energy dependent (see Energy Response Curve )
Photomultiplier Tube
3.8 cm (1.5 in) diameter
Operating Voltage
500 to 1200 V (typical)
Connector
Series "C"
Construction
Aluminum housing with beige powder coat
Size (D x L)
5.1 x 17.8 cm (2 x 7 in)
Weight
0.5 kg (1 lb)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item