Model 44-7
Ludlum Model 44-7, alpha-beta-gamma detector
- Manufacturer
- Ludlum Measurements
- Reference
- Model 44-7
- Product line
- Detector probes: alpha, beta and gamma
Specifications
- Product
- Ludlum Model 44-7, alpha-beta-gamma detector
- Ludlum part number
- 47-1536
- Indicated Use
- Alpha, beta, gamma survey; Sample counting
- Detector Type
- End window, halogen-quenched G-M
- Window
- 2.0 ± 0.3 mg/cm 2 mica
- Window Area
- Active: 6 cm 2 (0.93 in 2 ) Open: 5 cm 2 (0.78 in 2 )
- Efficiency (4π)
- 14 C: 2% 90 Sr/ 90 Y: 10% 99 Tc: 6% 239 Pu: 7% 125 I: 0.1%
- Typical Sensitivity ( 137 Cs gamma)
- 2100 cpm per mR/hr
- Exposure Energy Response
- Energy dependent
- H*(10) Energy Response
- Energy dependent
- Typical Background (10 µR/hr field)
- 40 cpm
- Typical Dead Time
- 200 μs
- Operating Voltage
- 900 V
- Connector
- Series "C" (others available)
- Construction
- Anodized aluminum housing
- Size (D x L)
- 4.6 x 14.7 cm (1.8 x 5.8 in)
- Weight
- 0.5 kg (1 lb)
Add this reference to my quote
Set the quantity and delivery on the interactive catalogue.
Request a quoteAll references for this itemOther references in this line
- Model 44-10Ludlum Model 44-10, gamma detector
- Model 44-11Ludlum Model 44-11, gamma detector
- Model 44-132Ludlum Model 44-132, gamma scintillator detector
- Model 44-159-1Ludlum Model 44-159-1, CsI scintillator detector
- Model 44-2Ludlum Model 44-2, gamma detector
- Model 44-20Ludlum Model 44-20, gamma detector
- Model 44-62Ludlum Model 44-62, gamma detector
- Model 43-1-1Ludlum Model 43-1-1, alpha-beta detector
- Model 43-135-1Ludlum Model 43-135-1, large area 2π proportional detector
- Model 43-143-1Ludlum Model 43-143-1, alpha-beta detector
- Model 43-147Ludlum Model 43-147, alpha-beta detector
- Model 43-2-2Ludlum Model 43-2-2, alpha-beta detector