Model 44-9-18

Ludlum Model 44-9-18, extended reach alpha-beta-gamma detector

Manufacturer
Ludlum Measurements
Reference
Model 44-9-18

Specifications

Product
Ludlum Model 44-9-18, extended reach alpha-beta-gamma detector
Ludlum part number
47-2940
Indicated Use
Alpha, beta, gamma survey; Sample counting
Detector Type
Pancake-type, halogen quenched G-M
Window
2.0 ± 0.3 mg/cm 2 mica, protected by a stainless steel screen (79% open)
Window Area
Active: 15.5 cm 2 (2.4 in 2 ) Open: 12.3 cm 2 (1.9 in 2 )
Efficiency (4π)
14 C: 5% 90 Sr/ 90 Y: 22% 99 Tc: 15% 32 P: 32% 239 Pu: 15% 99m Tc: ≤ 1% 125 I: 0.2%
Typical Sensitivity ( 137 Cs gamma)
3300 cpm per mR/hr
Energy Response
Energy dependent (see Energy Response Curve on Graphs tab)
H*(10) Energy Response (with optional filter 2002-1050)
Within 20% from 20 keV to 1.2 MeV
Typical Background (10 µR/hr field)
30-60 cpm
Typical Dead Time
80 μs
Operating Voltage
900 Vdc
Connector
Series "C" (others available)
Input Sensitivity
≥ -30 mV
Construction
Aluminum housing with beige powder-coat finish, stainless steel screen (79% open)
Size (H x W x L)
4.6 x 6.9 x 58.4 cm (1.8 x 2.7 x 23 in.) 0.9 m (36 in.) fully extended with gooseneck type flexible extension handle
Weight
0.7 kg (1.5 lb)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item