Model 12SA

Ludlum Model 12SA, microR survey meter with alarm

Manufacturer
Ludlum Measurements
Reference
Model 12SA

Specifications

Product
Ludlum Model 12SA, microR survey meter with alarm
Ludlum part number
48-2621
Indicated Use
Low-level (μR) gamma survey
Detector
NaI(Tl) scintillator, 2.5 x 2.5 cm (1 x 1 in.)
Typical Sensitivity ( 137 Cs)
175 cpm per µR/hr
Energy Response
Energy dependent
Alarm
Indicated by red lamp and audible tone
High Voltage
Adjustable from 200 to 1500 Vdc
Threshold
Adjustable from -6 to -32 mV
Linearity
±10% of true reading
Meter Dial
0 - 3 µR/hr, BAT TEST (Meter Face tab lists other available meter faces)
Controls
Rotary Selector Switch: off, battery check, range selections for x1, x10, x100, x1000 Reset Button: to zero meter High Voltage Check Button: view high voltage setting on meter face Audio Switch: on/off, built-in unimorph
Power
Two each "D" cell batteries (housed in externally accessible sealed compartment)
Battery Life
Exceeds 600 hours, may be checked on meter face
Construction
Cast and drawn aluminum with beige powder coat
Size (H x W x L)
22.2 x 21.6 x 8.6 cm (8.8 x 8.5 x 3.4 in.) with handle
Weight
2.1 kg (4.6 lb), including batteries
Meter Face
Image (click to view larger)

Add this reference to my quote

Set the quantity and delivery on the interactive catalogue.

Request a quoteAll references for this item