Ludlum Model 9-4RF, rf-immune air ion chamber survey meter

Manufacturer
Ludlum Measurements
Reference
Model 9-4RF

Specifications

Product
Ludlum Model 9-4RF, rf-immune air ion chamber survey meter
Ludlum part number
48-4338
Detection Range
0 to 500 mSv/h (0 to 50,000 mR/hr)
Energy Response
Within 20% of true value from 40 keV to 2 MeV
Linearity
Reading within 10% of true value
Response Time
Approximately 5 seconds for 90% of final meter deflection on the x1 and x10 scales, and 3 seconds on the x100, x1k and x10k scales
Chamber Volume
220 cm 3 (13.4 in 3 )
Chamber Construction
Carbon coated acrylic
Window
7 mg/cm 2 metalized polyester; with slide open, allows gamma detection to 6 keV
Window Area
40 cm 2 (6.2 in 2 ) (31.5 cm 2 [4.9 in 2 ] open with optional 79% open screen)
Side Wall
1000 mg/cm 2 aluminum and acrylic
Beta Shield
Retractable 1000 mg/cm 2 phenolic slide with side button control
Beta Response
Factor of 4.8 difference between window open and closed measurements when exposed to a uranium slab
Range Selection
Instrument Off, x10k, x1k, x100, x10, x1
Zero Adjust
One-turn potentiometer to zero reading
Battery
Pushbutton used to check battery capacity
Reset
The RESET circuit grounds the chamber and opens the chamber connection to the electrometer. This causes a hard upscale transient on the x1 scale. Hold the RESET down for 4 seconds to allow the transient to decay. When th
Display Light
Toggles a white LED display On/Off
Calibration
Digitally set via USB to serial computer interface, stored in non-volatile memory